DX NCPA Measurements

16:50 Finally get everything running.
  • Had troubles with telescope vibrations (my bad, running at 99 Hz not 990)
  • An error in a matrix inversion for Gopt. Alfio say it is a known bug and he will fix it soon
  • User error: loaded ModalDisturb instead of CmdDisturb
  • Gopt wouldnt work because loop was running at 99 Hz rather than 990 Hz (see above)
  • luci_ncpa.py would not request a luci cube. Would not even send ssh. Optical Loop Data process was down
  • flao could not write again in /towerdata/wfs_calib_W2/luci/backgrounds/bin1. chmod ugo+w bin1
  • Had trouble getting LUCI cubes. Went from failure 1 request in 10 to 1 request in 3. Tom restart scheduler

Rotator Angle = 30
Measurement Z4 Z5 Z6 Z7 Z8 Z9 Z10 Z11
23 April 2015 -26 124 88 -21 -5 -9 4 -9
First pass 29 -76 38 13 45 10 -11 -10

20:50 Finished RotAng = 0 Rotator Angle = 0 |||||||||
Measurement Z4 Z5 Z6 Z7 Z8 Z9 Z10 Z11
23 April 2015 -26 124 88 -21 -5 -9 4 -9
First pass 29 -60 -67 0 37 7 15 -10

Rotator Angle = 60
Measurement Z4 Z5 Z6 Z7 Z8 Z9 Z10 Z11
First Pass -31 16 90 36 25 0 -10 -10
-- %USERSIG{DougMiller - 2016-05-18}%
Topic revision: r3 - 18 May 2016, GregoryTaylor
This site is powered by FoswikiCopyright © by the contributing authors. All material on this collaboration platform is the property of the contributing authors.