DX NCPA Measurements

13:15 Closed loop

13:15 Start on rotator angle = 270

14:28 Troubles getting remote script to run on LUCI.
  • Turn off all three LUCI1 boxes (readout, instrument, mirror)
  • Turn off mount
  • finally work

14:39 Start collecting data at 270
  • start with z5/6 50/20 nm
  • start with

14:50 Troubles with finding LUCI background in Strehl calclation
  • Alfio fixed by changing directory permissions on /towerdata/wfs_calib_W2

15:00 Fixing bad actuator list found by Guido IT#4585
  • Flat still bad

17:09 Start RotAng=270
Rotator Angle = 270
Measurement Z4 Z5 Z6 Z7 Z8 Z9 Z10 Z11
23 April 2015 -26 50 63 -16 0 0 0 -9
First pass 29 30 44 -32 19 5 2 -10

20:25 Start RotAng=330
Rotator Angle = 330
Measurement Z4 Z5 Z6 Z7 Z8 Z9 Z10 Z11
23 April 2015 -26 26 -116 -9 24 7 10 -9
First pass 29 18 -104 -13 48 -13 10 -10

20:15 Finished RotAng=150
Rotator Angle = 150
Measurement Z4 Z5 Z6 Z7 Z8 Z9 Z10 Z11
23 April 2015 -26 14 -130 32 -9 13 7 -9
First pass 29 11 -70 24 -20 2 0 -10

20:25 Start RotAng=90
Rotator Angle = 90
Measurement Z4 Z5 Z6 Z7 Z8 Z9 Z10 Z11
23 April 2015 -26 124 88 -21 -5 -9 4 -9
First pass 29 95 37 40 10 -10 -10 -10

-- %USERSIG{DougMiller - 2016-05-17}%
Topic revision: r5 - 18 May 2016, DougMiller
This site is powered by FoswikiCopyright © by the contributing authors. All material on this collaboration platform is the property of the contributing authors.
Ideas, requests, problems regarding Foswiki? Send feedback