SX NCPA Measurements

05:13

Rotating to 180 degrees to begin re-checking NCPA

Loop params:

990Hz

binning: 1

3 lambda/D

05:27

Started NCPA with mode 5

06:02 remembered to sync wfs >:(

Mode 5: 67.5 [64.2,77.2,60.6,68.1] Original value 73, 5.5 nm difference, too small to measure, staying with original.

Mode 6: -26.4 [-27.3,-28.2,-19.1-30.8] Original value -28

Mode 7: 0 [-1.93,3.74,1.01,-2.96] Original value is 8

08:02

Had strange error and loop opened. Starting again.

Mode 8: -3 [-2.7,-3.17,-3.07] Original value is 10

Mode 11: -12 [-16.4,-7.9] Original value is -13 (Loop kept opening)

09:31

Moved to 120 degrees, 1 measurement per mode.

Mode 5: -44.6 Original -40

Mode 6: -16 Original -15

Mode 7: -19 Original -10

Mode 8: 14.6 Original 4

Mode 9: 5 Original 7

-- %USERSIG{DougMiller - 2016-04-06}%