02:20 First AO Preset
  • Al's Asteroid
  • Sideral tracking on a non-sideral object
  • AO system is "guiding"
  • Bright so Bin 1
  • First StartAO failed because the auto gain for All went to high (~2.0) and the mirror safe-skipped
  • Modified ~/TEST/conf/wfs/current/processConf/autogain.conf
    • changed bin1_tt_max from 3.0 to 2.0.
    • we had change this value from 2.0 to 3.0 yesterday night to try to get the TT gain found by auto gain to be higher. Was finding 0.3 not 1.0 as we expected
  • After third dither pauseAO, the resumeAO failed
    • dither offsets were in absolute coords which means the offsets were from the original position rather than the new "guided" position
    • LUCI scripts with relative offsets should solve this problem.

03:30 CL bin 3
  • Gopt is not detecting the disturbance. No NCPA
  • 033131
  • Looks like good correction. Could maybe have higher gain
  • Bin 3 Modal Plot:
    bin3_033131.jpg

04:30 Loop open because of safe skip
  • Greg close the loop manually

04:31 Loop open, skip frame

04:43 Preset CL
  • After auto gain, lower all gains a bit
  • still seeing a bit to TT flashing lower all gains a bit
  • 044816

04:50 CL with StartAO
  • 045149

04:58 Clouds coming over the last 40 minutes
  • Changing flux on pyramid of ~3 mag
  • 045448

05:10 Switch to AO performance tests because of the variable sky opacity
  • Preset to AO306 (6.9 mag)
  • Clouds clear, too bright on LUCI

05:26 Preset to AO309
  • 8 mag
  • RunAO failed: timeout on baysidey

05:36 Preset to AO309
  • 9.0 mag on pyramid
  • rotator 306
  • NCPA ON
    • 054603
    • 054635
    • 054709
    • bad background

05:45 Take new BG

05:54
  • 9.1 mag
  • NCPA ON
    • 055435
    • 055514
    • 055544
  • NCPA OFF
    • 055622
    • 055832
    • 055903

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20160320_055435 ??? 9.3 266561 5 1.02 0.48 184300 1 400 990 0.5 0.2 0.2 3 114 ON 49.7 FeII clear 8579 1 ONSKY zero.fits - long stop of 80 frames at frame 1478
20160320_055514 ??? 9.3 266093 4 1.14 0.48 184300 1 400 990 0.5 0.2 0.2 3 111 ON 47.4 FeII clear 8579 1 ONSKY zero.fits - long stop of 69 frames at frame 2644 - Background correction failed
20160320_055544 ??? 9.3 265747 5 1.49 0.55 184300 1 400 990 0.5 0.2 0.2 3 112 ON 45.4 FeII clear 8579 1 ONSKY zero.fits - long stop of 60 frames at frame 3586
20160320_055622 ??? 9.3 265293 5 1.46 0.50 184300 1 400 990 0.5 0.2 0.2 3 114 ON 43.7 FeII clear 8579 1 ONSKY zero.fits - long stop of 65 frames at frame 1028 - Background correction failed
20160320_055832 ??? 9.3 263734 5 1.14 0.46 184300 1 400 990 0.5 0.2 0.2 3 113 ON 46.3 FeII clear 8579 1 ONSKY zero.fits  
20160320_055903 ??? 9.3 263368 3 1.27 0.59 184300 1 400 990 0.5 0.2 0.2 3 116 ON 46.8 FeII clear 8579 1 ONSKY zero.fits - long stop of 65 frames at frame 1659
IDL string: ['20160320_055435','20160320_055514','20160320_055544','20160320_055622','20160320_055832','20160320_055903']
SR 0 (NaN)
SEEING (WFS) 0.510935 (0.04762)
SEEING (DIMM) 1.25441 (0.191481)

  • Modal Plot 055435:
    modal_055435.jpg

  • PSF 055435:
    psf_055435.jpg

05:59 200 frames
  • 055934

06:11 Preset
IDL> iif_runao
===> Error in irc_runao
        err = Resmsg[0]  = RunAO result status: Error
        err = Resmsg[1]  = AcquireRef{FLAO}[206] -  dx:0 dy:0 slNull: f1:Filter #0 f2:Filter #0 freq:990 mag:9.498 nBins:1 nModes:400 r0:0 snMode:0 strehl:0 ttMod:3
        err = Resmsg[2]  = StartAO{FLAO}[207] - IDL variable undefined 'gainv'
        err = Resmsg[3]  = StartAO{FLAO}[207] - IDL variable undefined 'gainv'

06:20 CL at 175
  • NCPA ON
  • Seeing 1.5-2.4
  • Can close loop, but big vibrations
  • Handover to Alexander for Active Flexure Compensation

08:55 LUCI Preset
Sun Mar 20 08:58:44.350 2016 left RunAO alert for LUCIFER: AcquireRef{FLAO}[251] - Timeout error on baysidex (-5001) TIMEOUT_ERROR
Sun Mar 20 08:58:44.361 2016 left RunAO command failed from LUCIFER, type NORMAL, leftRePoint 0, rightRePoint 0

09:02 LUCI Preset successful
  • Reference star is a 0.15" binary
  • bin 2
  • 091503

09:20 LUCI Preset
  • failed, CCD39 stopped, but restarted itself
Sun Mar 20 09:20:14.689 2016 left Preset alert for LUCIFER mode ADAPTIVEACE_ACTIVE: PresetAO{FLAO}[282] - Reading shared buffer masterdiagnostic.L:OPTLOOPBUF (-5001) TIMEOUT_ERROR
Sun Mar 20 09:20:14.705 2016 left Preset command failed for LUCIFER mode ADAPTIVEACE_ACTIVE with guidestars, offset, AO reference, Sound=presetFailed.wav

09:28 LUCI Preset
  • Failed. Looked like bayside y stage not moving
Sun Mar 20 09:27:39.488 2016 left RunAO alert for LUCIFER: AcquireRef{FLAO}[285] - Command timeout - -13107: Arbitrator interface generic error
Sun Mar 20 09:27:39.497 2016 left RunAO command failed from LUCIFER, type NORMAL, leftRePoint 0, rightRePoint 0

09:33 LUCI Preset
  • Failed
Sun Mar 20 09:33:48.971 2016 left RunAO alert for LUCIFER: AcquireRef{FLAO}[291] - Timeout error on baysidex (-5001) TIMEOUT_ERROR
Sun Mar 20 09:33:48.980 2016 left RunAO command failed from LUCIFER, type NORMAL, leftRePoint 0, rightRePoint 0

09:45 Preset Success
  • 094326
  • Modal Plot 091503:
    modal_091503.jpg

10:08
  • 100812
  • Modal Plot 100812:
    modal_100812.jpg

10:47 Preset from IDL
  • Failed
Sun Mar 20 10:46:33.553 2016 left RunAO alert for LUCIFER: AcquireRef{FLAO}[398] - Timeout error on baysidex (-5001) TIMEOUT_ERROR
Sun Mar 20 10:46:33.564 2016 left RunAO command failed from LUCIFER, type NORMAL, leftRePoint 0, rightRePoint 0

10:51 Preset from IDL
  • Failed
Sun Mar 20 10:46:33.553 2016 left RunAO alert for LUCIFER: AcquireRef{FLAO}[398] - Timeout error on baysidex (-5001) TIMEOUT_ERROR
Sun Mar 20 10:46:33.564 2016 left RunAO command failed from LUCIFER, type NORMAL, leftRePoint 0, rightRePoint 0

11:08 CL on 9.7 mag start. Oops, Ks filter
  • NCPA ON
    • 110935
    • 111029
    • 111152
  • NCPA OFF
    • 111228
    • 111254
    • 111319

11:20 Took 50 and 200 darks

11:21 FeII
  • rot=-5
  • NCPA = 30 nm PtoV
  • NCPA ON
    • 112951
    • 113053
    • 113135
  • NCPA OFF
    • 113238
    • 113303
    • 113328

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20160320_112951 ??? 9.5 277687 2 NaN 0.25 184300 1 400 990 0.6 0.2 0.1 3 93 ON 55.1 FeII clear 172 50 ONSKY zero.fits - long stop of 88 frames at frame 2749
20160320_113053 ??? 9.5 277805 2 NaN 0.28 184300 1 400 990 0.6 0.2 0.1 3 93 ON 57.4 FeII clear 172 50 ONSKY zero.fits - long stop of 62 frames at frame 2257
20160320_113135 ??? 9.5 277878 2 NaN 0.31 184300 1 400 990 0.6 0.2 0.1 3 92 ON 55.5 FeII clear 172 50 ONSKY zero.fits - long stop of 98 frames at frame 1523
20160320_113203 ??? 9.5 277924 2 NaN 0.26 184300 1 400 990 0.6 0.2 0.1 3 96 ON 54.7 FeII clear 172 50 ONSKY zero.fits  
20160320_113238 ??? 9.5 277980 2 0.79 0.25 184300 1 400 990 0.6 0.2 0.1 3 97 ON 55.3 FeII clear 172 50 ONSKY zero.fits  
20160320_113303 ??? 9.5 278016 2 0.74 0.26 184300 1 400 990 0.6 0.2 0.1 3 94 ON 52.4 FeII clear 172 50 ONSKY zero.fits - long stop of 83 frames at frame 1688
20160320_113328 ??? 9.5 278051 2 0.80 0.28 184300 1 400 990 0.6 0.2 0.1 3 96 ON 52.6 FeII clear 172 50 ONSKY zero.fits - long stop of 53 frames at frame 2160
IDL string: ['20160320_112951','20160320_113053','20160320_113135','20160320_113203','20160320_113238','20160320_113303','20160320_113328']
SR 0 (NaN)
SEEING (WFS) 0.270149 (0.0199598)
SEEING (DIMM) 0.79787 (NaN)

11:45 FeII
  • Rot = 225
  • NCPA = 150 nm PtoV
  • NCPA OFF
    • 114458
    • 114550
    • 114623
  • NCPA ON
    • 114709
    • 114742
    • 114806
    • 114837 200 frames

TrackNo RefStar Mag El Wind DIMM/OL Rec bin #mod freq gain mod nph AntiDrift SR filter exp #frames disturb SN notes
20160320_114458 ??? 9.5 278309 1 1.03 0.41 184300 1 400 990 0.6 0.2 0.1 3 91 ON 47.5 FeII clear 172 50 ONSKY zero.fits  
20160320_114550 ??? 9.6 278272 1 0.96 0.38 184300 1 400 990 0.6 0.2 0.1 3 88 ON 46.7 FeII clear 172 50 ONSKY zero.fits - long stop of 105 frames at frame 399
20160320_114623 ??? 9.6 278243 2 0.80 0.35 184300 1 400 990 0.5 0.2 0.2 3 90 ON 47.5 FeII clear 172 50 ONSKY zero.fits - long stop of 63 frames at frame 3722
20160320_114709 ??? 9.6 278198 1 0.98 0.36 184300 1 400 990 0.5 0.2 0.2 3 85 ON 55.5 FeII clear 172 50 ONSKY zero.fits  
20160320_114742 ??? 9.6 278163 1 0.93 0.36 184300 1 400 990 0.5 0.2 0.2 3 90 ON 52.6 FeII clear 172 50 ONSKY zero.fits  
20160320_114806 ??? 9.6 278135 2 1.07 0.32 184300 1 400 990 0.5 0.2 0.2 3 87 ON 53.5 FeII clear 172 50 ONSKY zero.fits - long stop of 65 frames at frame 2337
20160320_114837 ??? 9.6 278096 2 0.98 0.33 184300 1 400 990 0.5 0.2 0.2 3 87 ON 49.4 FeII clear 172 200 ONSKY zero.fits - long stop of 59 frames at frame 3757
IDL string: ['20160320_114458','20160320_114550','20160320_114623','20160320_114709','20160320_114742','20160320_114806','20160320_114837']
SR 0 (NaN)
SEEING (WFS) 0.358038 (0.0299365)
SEEING (DIMM) 0.964558 (0.0860518)

  • Modal Plot 114806:
    modal_114806.jpg

  • PSF 114806:
    psf_114806.jpg

-- %USERSIG{DougMiller - 2016-03-19}%
I Attachment Action Size Date Who Comment
bin3_033131.jpgjpg bin3_033131.jpg manage 86 K 20 Mar 2016 - 03:50 DougMiller Bin 3 Modal Plot
bin3_033131_2.jpgjpg bin3_033131_2.jpg manage 46 K 20 Mar 2016 - 03:50 DougMiller bin 3 Modal Plot #2
modal_055435.jpgjpg modal_055435.jpg manage 104 K 20 Mar 2016 - 06:20 DougMiller Modal Plot 055435
modal_091503.jpgjpg modal_091503.jpg manage 99 K 20 Mar 2016 - 10:01 DougMiller Modal Plot 091503
modal_100812.jpgjpg modal_100812.jpg manage 100 K 20 Mar 2016 - 10:11 DougMiller Modal Plot 100812
modal_114806.jpgjpg modal_114806.jpg manage 101 K 20 Mar 2016 - 11:57 DougMiller Modal Plot 114806
psf_055435.jpgjpg psf_055435.jpg manage 92 K 20 Mar 2016 - 06:21 DougMiller PSF 055435
psf_114806.jpgjpg psf_114806.jpg manage 89 K 20 Mar 2016 - 11:58 DougMiller PSF 114806
Topic revision: r5 - 20 Mar 2016, DougMiller
This site is powered by FoswikiCopyright © by the contributing authors. All material on this collaboration platform is the property of the contributing authors.
Ideas, requests, problems regarding Foswiki? Send feedback