SX FLAO Reconstructor Measurements

Summary

TO DO: - create 10 mode (or 21 mode) recons for ASM - create 10 mode recons for low order mode closure during AcquireRefAO - analyze bin 4, 48 iteration set - analyze bin 1, slave KL mode set - remove extra pupil files - create ACE and TTM tables

Details

03:29 RR is install, at Zenith, Stow Pins in, light minimal

03:20 Getting ready
  • Power on wfssx
  • Power on flowerpot, cube stage, cube rotator, lamp
  • Choose board setup 20160217_RR the Greg saved yesterday
    • failed waiting for cuberot
    • homed cuberot
    • success
  • Turn Lamp to 100%
  • Change FW1 from Dichroic 600-1000 to Silver Mirror
  • Change FW2 from 900-10 to Empty
  • Rotated cube to 92.4. Could see internal source
  • Set SX Adsec Flat
  • Authorize RETROREF@LFBG and NONE
  • Collimate SX M2
  • Rotated cube to 182.4. Could not see RR return. Could not set internal reflection on right of TV
  • Start hexa_spiral. When focus changed could see the return beam
    • Could have just change focus to 6.5, found the star and adjusted RX, RX
  • Put star on TV hotspot by moving bayside stages
  • Set loop parameters 1, 600, 3.0

04:10 Close loop and CCD39 goes to Not Live
  • Clicked FLAO on WFS Control GUI Not Live
  • Powered off Little Joe CCD39 Not Live
  • Stop and started Slope computer and CCD 39 Controller processes Not Live
  • power off wfs, stop/start processes. Live.

04:54 Closed loop, 400 modes
  • Opened loop and check cube rotator = 182.028

Bin 1

05:00 Start taking new Recons

05:00 Distrub
  • 400 modes
  • 4 push push
  • 1 cycle
  • 170124

05:02 Focus on AdSec
  • closed loop to offload focus

05:05 Acquire
  • 4 iter
  • gain0.05
  • recon 200644 (pupil correct)
  • bin 1
  • 3 lam/D
  • 600 Hz
  • 170739

05:11 Analyze
  • 170938

05:13 Acquire
  • 48 iter
  • gain0.05
  • recon 170938 (pupil 20160209-190618)
  • bin 1
  • 3 lam/D
  • 600 Hz
  • 171425

05:35 Analyze
  • 172957 (GOOD RECON TO USE)

05:42 Atmosphere 1.2" seeing 600 Hz 3 lam/D
  • dist_flao1_20160217_atm_s1.2_L040.0_v15.0_ovfreq600.00_sd3892.fits
  • 1.3, 0.8, 0
  • 174616

05:59 atmosphere 0.8" seeing, 1000Hz, 3 lam/D
  • dist_flao2_20111210_atm_s0.8_L040.0_v15.0_ovfreq1000.00_sd1983.fits
  • 0.6, 0.5, 0.2
  • 180314
  • Optimize Gain
    • 0.28, 0.58, 0.2
    • 180428

06:15 New flat: 20160218_flat2.sav

Bin 2

06:17 Bin 2
  • tried to close the loop with yesterday's 153 modes (221346) but unstable
  • closed loop with 10 mode recon (203528)

06:23 Disturb
  • bin 2
  • 66 modes
  • 4 frame
  • 7 cycles
  • optomize_amono_3lambda.fits
  • 182308

06:25 Acquire
  • 4 iter
  • 66 modes
  • 203528
  • gain0.05
  • 182554

06:28 Analyze
  • 66 modes *182735

06:29 Disturb
  • bin 2
  • 153 modes
  • 4 frame
  • 3 cycles
  • optomize_amono_3lambda.fits
  • 183030

06:31 Acquire
  • 4 iterations
  • 153 modes
  • 183436

06:38 Analyze
  • 183848

06:40 Acquire
  • 48 interation
  • 153 modes
  • bin 2
  • 600 Hz
  • 3 lam/D
  • 184205

06:56 Analyze
  • 185624 (USE THIS BIN2 RECON)

07:10 atmosphere 0.8" seeing, 990Hz, 3 lam/D, bin 2
  • dist_flao2_20111210_atm_s0.8_L040.0_v15.0_ovfreq1000.00_sd1983.fits
  • 185624
  • manual gain 0.7, 0.7
    • 191226
  • auto gain 0.38, 0.55, 0.55
    • 191432

Bin 4

07:16 Close loop with yesterday's recon
  • 225822
  • 6 lam/D
  • 600 Hz
  • Bin 4

07:22 Disturb
  • optimized_amano_6l.fits
  • 4 frame
  • 7 cycles
  • 66 modes
  • 192240

07:23 Acquire
  • 192452

07:28 Analyze
  • bin 4
  • iter 4
  • 192635

07:50 Disturb
  • 66 modes
  • optimized_amano_6ldiv2.fits
  • 194911

07:51 Acquire
  • use Rec 225822
  • 4 iter
  • 195200

07:53 Analyze
  • 195401
  • bad SN plot and single values

08:11 Disturb
  • optomized_amano_6ldiv1.5.fits
  • 201100

08:14 Acquire
  • disturb with div1.5
  • 201304

08:17 Disturb
  • bin 4
  • 66 modes
  • 600_optimazed_6l
  • 201630

08:19 Acquire
  • big bump in the lower left side of the position, current and command display

08:21 Disturb
  • optimized_amano_6l (again)
  • 202007

08:22 Acquire
  • 202217

08:23 Analyze
  • 202355
  • not stable. Line in command display

08:29 Test
  • 225822 from yesterday
  • opt. loop data 202928

08:35
  • analyze yesterday's bin4 set 22511?
  • 203324

09:11 Disturb
  • optimized_amano_3l
  • 4
  • 13
  • 211211

09:15 Acquire
  • 211555

Analyze
  • 211739

09:30 Distrub
  • optimaized_amano_6l.fits
  • 212825

09:35 Acquire
  • 4 iter
  • 213053

09:40 Analyze
  • 213352

09:44 Acquire
  • 48 iter
  • bin 4
  • 6 lam/D
  • 600 Hz
  • 214223

  • 220323 recon

Bin 1 test

10:00 Closed loop
  • 164809

10:05 Manual gain
  • .25, .45, 0.2
  • 221241
  • .25, .45, .25
  • 221209 *
  • .25, .35, .30
  • 221112

10:15 Auto gain
  • .28, .40, .19
  • 221421 *
  • .28, .40, .24
  • 221501
  • .28, .40, .29
  • 221538
  • .28, .50, .24
  • 221834

  • 164809

Slave KL_v20

10:36 Disturb
  • KL_v20
  • 4 pp
  • 1 cycle
  • opt_amano_3l
  • 223622

10:37 Acquire
  • bin 1
  • 600 Hz
  • 3 lam/D
  • 48 iter
  • disturb 223622
  • 223945

10:55 Analyze
  • 225429 (SLAVE INNER ACTUATORS)

Test Slave KL_v20

11:04 Closed loop with 225429
  • .3, .4, .2
  • 230416
  • .3, .4, .25
  • 230535
  • .3, .4, .3
  • 230644

Bin1, no slave, 350, 300, 250 modes

11:12 Analyze
  • intmat 171425
  • 350 modes
  • 231106

11:14 Test
  • .7, .4, .2
  • 231602
  • .7, .4, .3
  • 231637

11:18 Analyze
  • intmat 171425
  • 300 modes
  • 231747

11:20 Test
  • .7, .4, .25
  • 232147 .7, .4, .35
  • 232229

11:23 Analyze
  • intmat 171425
  • 250 modes
  • 232326

11:28 Test
  • .7, .4, .3
  • 232905
  • .7, .4, .4
  • 232950

-- %USERSIG{DougMiller - 2016-02-18}%