LUCI2/AO Commissioning 20150601 UT

Observers: DMiller, GBrusa, JHill, SW Group (Tucson)
Telescope Operator: DHuerta (LBTO)
AO Support: DMiller (LBTO)
Software Support: (Tucson)

Summary

Details

06:15 WFS Arb was off, so power up from AOS failed, after full timeout. AOS should know no communication and fail immediately.

06:20 Power on WFS

06:30 Steve preset to 8.3 (see 9.2) and close loop

06:40 Jochen preset to 8.1 ref star and close loop

06:54 CCD39 went offline on a PauseAO request.
  • Tried to set binning to 2x2 (from 1x10) Did not change from Not Live
  • Stopped/started CCD39 process and set to 2x2 binning. Went to state Configuring, but then stopped Not Live
  • Stopped/started CCD39 and Slope Computer Ctrl. Sent PresetAO and worked

07:09 NCPA installed -26, -90, -90, -16, 16, 0, 0, -10

07:51 Skip Frames and loop open on an offset

09:29 Loop Open, shell RIP and Recover Fail
  • fastdiagn.R |WAR| 9057193|2015-06-01 09:29:35.958863| MAIN > Timeout - Will emergencyReact next loop [FastDiagnostic.cpp:1097]
  • fastdiagn.R |WAR| 9057194|2015-06-01 09:29:37.959720| MAIN > Timeout - Will emergencyReact now [FastDiagnostic.cpp:1092]
  • fastdiagn.R |ERR| 9057195|2015-06-01 09:29:37.959761| FUNCTEMERGENCYST > emergencyReact = 1 [Funct.cpp:94]

09:49 Had offset from reference star in center of field to ~2' (in correct direction) in step of 15". The offset from ~2' to 0 failed, but the TCS and LUCI were not informed properly
  • WfsArbitrator_1557 |ERR| 1967|2015-06-01 09:49:17.336899| MAIN > RETRY: Offset outside FoV limits. (-5002) VALUE_OUT_OF_RANGE_ERROR
  • WfsArbitrator_1557 |ERR| 1968|2015-06-01 09:49:17.336925| MAIN > Traceback (most recent call last):
  • WfsArbitrator_1557 |ERR| 1969|2015-06-01 09:49:17.336945| MAIN > File "/home/aoacct/AO/current/lib/python/AdOpt/WfsArbScripts.py", line 2708, in arbFunc
  • WfsArbitrator_1557 |ERR| 1970|2015-06-01 09:49:17.336964| MAIN > errstr = func(*args)
  • WfsArbitrator_1557 |ERR| 1971|2015-06-01 09:49:17.336984| MAIN > File "/home/aoacct/AO/current/lib/python/AdOpt/WfsArbScripts.py", line 1846, in offsetXYinternal
  • WfsArbitrator_1557 |ERR| 1972|2015-06-01 09:49:17.337002| MAIN > raise AORetryException( e.code, errstr)
  • WfsArbitrator_1557 |ERR| 1973|2015-06-01 09:49:17.337021| MAIN > AORetryException: Offset outside FoV limits. - code: -5002 VALUE_OUT_OF_RANGE_ERROR

10:46 Ref star = 1.53 (pyramid), 11.2 (preset)
  • try to change New Freq to be able to make NCPA corrections
  • Changing New Freq seems to make no difference.
  • Will observer without NCPA

10:50 Safe Skip, Loop opens

11:00 Loop closed

-- %USERSIG{DougMiller - 2015-06-01}%