NCPA tests with LUCI2

00:50 Startup all systems

01:35 Star found on Acquisition camera
  • rotator at 150

01:42 N30 camera. Adjust M2 Hex and stages
  • -71.5, -56 (-4, -6)
  • Pupil look really good for 10 mode reconstructor

01:54 Acquire ref failed, camera lens not
  • Thu Apr 23 13:52:51.714 2015 aos.AcquireRef{FLAO}.failed - AcquireRef{FLAO}[81] - Camera lens centering did not converge after 180 seconds

1:54 presetao
  • Thu Apr 23 13:54:09.343 2015 aos.PresetAO{FLAO}.failed - PresetAO{FLAO}[82] - Reading shared buffer masterdiagnostic.R:OPTLOOPBUF (-5001) TIMEOUT_ERROR

01:55 retry presetao OK

01:56 acquireRefAO: Success. StartAO:

02:01 Closed loop

02:26 Get subarray set. Had a bit of trouble with array. Had selected "Show all RO information' on the display GUI and we saw 128x2048 (hardware read). We selected "Show selected subarr" and we see 128x128

02:30 Collect 20 background frames.
  • Save set to normal. NO
  • Set save to cube
  • collect background

02:40 Adjust Z5, z6 manually and find about (0, -100)

02:41 start NCPA acquisition at 150 rot angle
restart[4]  =  -26 # 59.7
restart[5]  =  14 #-35.0  # 0.9 # 3.5
restart[6]  =  -130 #30 # 30.0 # 16.0
restart[7]  =  32 # -30 # -12.0 # -17 ???
restart[8]  =  -9 #-30 # 2.0 # -11.0
restart[9]  =  13 #15 #  4.0
restart[10] =  7 #0 # -7.5
restart[11] =  -9 #-36 # -12

03:50 Finish 150 data collection

03:55 Stop loop, N3.75 camera, rotate to 90, move stages to -78, -55

05:05 Loop closed and positioned on subarry

05:05 Looks like Z5=100, z6=90
restart[4]  =  -30 # 59.7
restart[5]  =  124 #-35.0  # 0.9 # 3.5
restart[6]  =  88 #30 # 30.0 # 16.0
restart[7]  =  -21 # -30 # -12.0 # -17 ???
restart[8]  =  -5 #-30 # 2.0 # -11.0
restart[9]  =  -9 #15 #  4.0
restart[10] =  4 #0 # -7.5
restart[11] =  -9 #-36 # -12

06:13 Finish data collection at 90

06:40 start data collection at 30
  • looks like Z5=-50, Z6=50 to start

restart[4] = -30 # 59.7 restart[5] = -100 #-35.0 # 0.9 # 3.5 restart[6] = 70 #30 # 30.0 # 16.0 restart[7] = -2 # -30 # -12.0 # -17 ??? restart[8] = 14 #-30 # 2.0 # -11.0 restart[9] = 2 #15 # 4.0 restart[10] = -2 #0 # -7.5 restart[11] = -5 #-36 # -12

07:20 Finish

07:25 Rotate to 210

07:45 During "Gain Optimizaion" BCU47 stop communicating. Status Check on HW GUI is red X

07:47 PresetAO fail. no CCD47
  • Thu Apr 23 19:48:49.459 2015 aos.PresetAO{FLAO}.failed - PresetAO{FLAO}[110] - Command timeout - -13107: Arbitrator interface generic error

07:49 Rotate from 210 to 230. No change

07:50 Rotate from 230 to 300. No Change

07:56 Cycle power on BCU47 prog signal and Tip-Tilt

07:57 Rotate from 210 to 0. No change

07:58 Power off from AOSGUI. Fail
  • Thu Apr 23 19:58:24.380 2015 aos.WfsOff.failed - WfsOff[111] - thGetVar powerboard.R.fans_LJ.CUR error(-3606): No connection with MsgD

07:59 Cycle WFS Arbitrator process and Power off works.

08:02 Power on at AOSGUI. Fail
  • Thu Apr 23 20:02:46.570 2015 aos.WfsOn.failed - WfsOn[113] - Timeout waiting for pinger.R.bcu47 (-5001) TIMEOUT_ERROR

08:06 Power on at AOSGUI. Fail
  • Thu Apr 23 20:06:25.930 2015 aos.WfsOn.failed - WfsOn[114] - Timeout waiting for pinger.R.bcu47 (-5001) TIMEOUT_ERROR

08:07 Cycle "Box 47 Power Supply" on HW gui and BCU47 comes on (green box) for about 30 seconds. Then off (red box)

08:09 Cycle "Box 47 Power Supply" on HW gui and BCU47 does NOT come on

08:10 Rotator from 0 to 180. No Change

08:11 Turn off all power on HW gui

08:20 Power on from AOSGUI. BCU47 running as well as Tip-tilt. On for a few minutes. Changed board setup to calibUnitNight and BCU47 and Tip-Tilt power off

08:35 Power all off, power all on. BCU47 and T-T on for 10 seconds

08:40 Power off

08:41 Power on. BCU47 did not power on at all.

08:50 Done for today

08:56 change LUCIFER.sav sym link from 20150325_400modes.sav to 20131021_600modes_4d.sav. Ready for on-sky

-- %USERSIG{DougMiller - 2015-04-23}%
Topic revision: r3 - 23 Apr 2015, DougMiller
This site is powered by FoswikiCopyright © by the contributing authors. All material on this collaboration platform is the property of the contributing authors.