NCPA tests with LUCI2/AO

03:35 Connect to Jeff and Jenny on the mountain. Telescope is configured for LUCI2

03:45 DX w_stop, w_start

03:50 Start WFS from AOS GUI

03:52 Authorize LUCI on DX, NONE on SX

03:55 Connect to Alfio via Skype

04:08 Test image collection to LUCI2
  • Open Optical loop data window from wfseng window
  • click Slopes and set to some value of frames (100)
  • click save Pisces pdf: cube of __ , say 20
  • Start
  • Lower box says "Frames saved: 100 Speed, 66 Hz
  • Upper box says "Saved: tracking_number"
    • text is greyed out while data is collected.
    • Turns green when complete
    • check if connection to LUCI2 with "ps -eaf | grep ssh"
  • 20150417_160816
  • Alfio confirm image there. Full Frame

04:18 Reconfigure board setup
  • On HW GUI click "Board Setup" on the left
  • Click 02 Mar 15 - calibUnitDay
  • Takes about 2 minutes
  • WFS Control GUI does not give current board config
  • Flower pot should be on

04:20 Tom Setup LUCI2 in subwindow mode
  • 20150417_162408
  • Alfio confirm

04:27 Collect "Pisces Dark" on WFS Control GUI
  • 20150417_162717
  • Alfio confirm

04:29 Tom turn off sub window

04:35 Gustavo start to deploy ARGOS swingarm

04:37 AOeng@adsecdx % enable_rr_mode.sh
  • Yellow "Retroreflector installed" on AOSGUI

04:40 Setup hexapod
  • http://aowiki.arcetri.astro.it/FLAO/ConFigurations
  • At the bottom of the table "HEXAPOD POSITION RETROREFLECTOR"
  • 04/04/2015 -1.1 -3.6 4.3 -530 272 0 Argos calibration source
  • On PSFGUI right
    • disable Lookup Table and Temperature Correction
    • Enter Global Offset parameters and Update.
    • The Total Collimation values should be the Global Offsets

04:52 Tom put in clear filter and FEII

04:55 Tom put LUCI2 in Endless Mode
  • See a lot of light
  • see image

06:10 Gustavo find that the SX and DX AGROS light sources were swapped during the last ARGOS commissioning run
  • see image

06:24 Tom put in N30 camera in.
  • see image N30_1

06:35 best focus, with coma
  • see image N30_2

06:40 Collimate tertiary.

07:00 put source on the middle of LUCI2 with -1.1, -4.4, 4.6, -230, 230
  • WUnit stage at -67, -71.
  • still quite aberrated

07:00 Close down
  • Gustavo turn off ARGOS source and retract arm
  • Tom put in BLIND+BLIND and shutdown LUCI2
  • Doug rest adsec, power off WFS, disable_rr_mode

-- %USERSIG{DougMiller - 2015-04-17}%
Topic revision: r2 - 17 Apr 2015, DougMiller
This site is powered by FoswikiCopyright © by the contributing authors. All material on this collaboration platform is the property of the contributing authors.