Known history

When Where Who What
20130215 OAA JZ, GOX Delivered
20131011 OAA GOX Upgrade of the electronic panel on the external side
20131202 OAA GOX Repumped and ??
20140114 OAA MB, JZ, TM Removed from LgswUnitDX
20140130 OAA GOX Ship to MPE.
20150209 OAA MB, GOX Ship to INAF, installed to SX.
20151017 LBT GOX Update software and aligned Kontron to WFS camera #2 (see instructions here)
20150308 LBT GOX, TM, GR pumping camera
20150511 LBT TM, MB pumping camera
20150515 LBT TM, SR pumping camera
20161021 LBT TM Pressure read 30 mbar
20170501 LBT MR, MB This camera is leaking, It is now routinely pumped every day
20170712 LBT HG Perfomed leak check and found leak in the housekeeping connector. Applied Apiezon Q sealant around it with good results.
20180210 LBT GR Pressure was 12 mbar. Pumped down to 0.5 mbar. Last reading was 0.62 mbar
20181129 LBT GR Pressure was 54 mbar. Pumped down to 0.45 mbar (15 minutes). 5 minutes after closing the valve the reading was 0.55 mbar
20190815 LBT GR Pressure was 84 mbar. Pumped down to 0.58 mbar (10 minutes). 5 minutes after closing the valve the reading was 0.94 mbar
20200306 LBT GR, CB Pressure was 49-54 mbar (reading was changing). Pumped for 1 hour down to 0.43 mbar
20200903 LBT CB Pressure was >80 mbar. Pumped for 1 hour down to 0.41 mbar
20210120 LBT CB Pressure was 45-50 mbar. Pumped for 1 hour.
20210624 LBT GR Pressure was 67 mbar. Pumped for 2 hours down to 0.45 mbar
20210928 LBT CB Pressure was >50 mbar. Pumped down to 0.69 mbar (had to stop for rain needing telescope at horizon)
20220216 LBT CB Starting pressure was 42 mbar. Pumped down to 5.7 x10 -1
20220622 LBT CB Pressure at 43mbar.Down to 0.2mbar

Known issues

When What
20130215 Pressure sensor is miscalibrated (?). Typically stays at 0.2mbar and shows no variations at all
20130819 Pressure too high. Sensor passes from 0.2mbar to 7000mbar
20130828 Pumped. Diagnostic is now ok. Camera is apparently working ok
20131205 Horizontal Stripes effect began. See emails
20131223 CCD cannot be powered up. A current value is too high. See emals
20131230 A bent pin in the sequencer cable is found. See emails
20150304 ErrorCamera at startup: ThingsThatIDontUnderstand0010
20170501 The camera leak progressively got worse and camera needs to be pumped more often. In May17 run (DiaryRun16) the camera was pumped every day

Pressure sensor

The pressure sensor of this camera appears to be miscalibrated. Its floor value is 0.2mbar and it appears quite unsensitive to all changes. It passed suddenly to 7 bar!. It appears not linear at all.

When Where Pressure from pressure gauge

Pressure camera sensor

(Voltage reading, converted in mbar)

Humidity raw value

(water cooling only, at -20C, max value when warming up)

Comments
20151017 LBT 4 mbar - - -
20150308 LBT 54 mbar 49.5 mbar - Maintenance, camera pumped down to ~1mbar
20150313(?) LBT ~25 ~10 mbar (?) - Maintenance, camera pumped down to ~1mbar for ~30min
20150313(?)+1 LBT ~5 mbar   -  
20150511 LBT ??   - Maintenance, camera pumped down to ~1mbar
20150516 LBT 12 mbar 7.644 mbar - Maintenance, camera pumped down to ~1mbar
20180210 LBT 12 mbar - - Maintenance, camera pumped down to 0.5 mbar
20190815 LBT 84 mbar - - Maintenance, camera pumped down to 0.58 mbar
20200306 LBT 49-54 mbar (changing) 50 mbar - Maintenance, camera pumped down to 0.43 mbar (after 1 hour)
20200903 LBT >80 mbar 80 mbar 40% Maintenance, camera pumped down to 0.41 mbar (after 1 hour)
20210120 LBT ? 47 mbar 75% Maintenance, camera pumped for 1 hour.
20210624 LBT 67 mbar ? 45% Maintenance, camera pumped for 2 hours to 0.45 mbar
20210928 LBT ? >50 mbar 90% Maintenance, camera pumped to 0.69 mbar
20220216 LBT 42 mbar ? 55% Maintenance, camera pumped to 0.57 mbar
20220622 LBT   43mbar 50% Maintenance, camera pumped to 0.02 mbar

Maintenance

See WfsCamera

List of components devices

DeviceInfoForm edit

Name WFS Camera #1
SerialNumber 1
Location LgswUnitSx
MACAddress
DeviceType WfsCamera
Service Tag