PSF wobble on LUCI1

We close the loop on FLAO, we put in N30 mask and we spin the rotator w/ and w/o the TTOffload LUT:

23:15: the weather seems to clear, we try to open

NGS5466

00:00: ready to propagate...

  • DX
TN AO Gain AO CL TT Res. WF [nm rms] HO Res. WF [nm rms] AO rec DIMM ["] JCL PCL Notes
20170506_070559 1.5,0.4,0.3 T 127.941 171.643 20161205_222600 1.11273 T F
20170506_070947 1.5,0.4,0.3 T 146.464 128.896 20161205_222600 1.12909 T F
20170506_071211 2,0.4,0.3 T 156.573 121.875 20161205_222600 1.05612 T F
20170506_071239 2.5,0.4,0.3 T 137.85 128.934 20161205_222600 1.09224 T F
00:18: closing again...

SDSSJ1985

3:00: 1985 back to the sky -> <-

4:00: DX vib comp goes crazy again, we disable it...

4:16: DX BCU stacks...

TN AO Gain AO CL TT Res. WF [nm rms] HO Res. WF [nm rms] AO rec DIMM ["] JCL PCL Notes
20170506_102246 2.5,0.4,0.3 T 57.7634 119.489 20161205_222600 1.16333 T F
20170506_102628 2.5,0.4,0.3 T 57.8923 164.839 20161205_222600 1.11908 T F
20170506_102757 1.5,0.4,0.3 T 56.9937 167.055 20161205_222600 1.23703 T F
20170506_104626 1.5,0.4,0.3 T 59.1255 114.109 20161205_222600 1.24984 T F Auto
20170506_104841 1,0.4,0.3 F nan 534.171 20161205_222600 1.11375 T F Auto
20170506_105652 0.375, 0, 0 T 69.8479 591.892 20161205_222600 1.32538 T F Auto
20170506_105746 1.5,0.4,0.3 T 63.7429 133.841 20161205_222600 1.28845 T F
20170506_110506 0.075, 0, 0 F 138.086 536.744 20161205_222600 1.24583 T F Auto
20170506_110720 1,0.4,0.3 T 62.0519 126.699 20161205_222600 1.23222 T F Auto
20170506_110934 1,0.4,0.3 T 57.5277 118.95 20161205_222600 1.171 T F Auto
20170506_111147 0.1, 0, 0 T 86.5214 521.166 20161205_222600 1.15105 T F Auto
20170506_111400 1,0.4,0.3 T 58.2155 131.649 20161205_222600 1.22489 T F Auto

  • PBL when BCU stacks and LGSW is restarted the jitter loop in the controller is not disabled and the guider too. This causes the message:
2017-05-06 11:40:20.416225413 NOTICE PmToLM1Offload: Moving LM1 relative by {phi: 1.000'', theta: 1.000'', piston: 0.000 mm}
that really continuously moves LM1!
  • Camera lens loop on DX is very bad: can it left disabled if it is disabled before the offset?