03:15UT: doing flats.

SDSSJ1110+6459

03:45UT: loop closed.

TN AO Gain AO CL TT Res. WF [nm rms] HO Res. WF [nm rms] AO rec DIMM ["] JCL PCL Notes
20160521_034550 0.25,0.3,0.2 T 1002.29 134.277 20160312_032600 0.804729 T F
20160521_034610 0.25,0.3,0.2 T 1003.49 125.224 20160312_032600 0.794879 T F
20160521_034622 0.25,0.3,0.2 T 949.27 111.01 20160312_032600 0.780531 T F

04:03UT: mask aligned. Starting integration

TN AO Gain AO CL TT Res. WF [nm rms] HO Res. WF [nm rms] AO rec DIMM ["] JCL PCL Notes
20160521_040655 0.25,0.3,0.2 T 938.547 128.775 20160312_032600 0.792785 T F
20160521_040707 0.25,0.3,0.2 T 972.013 128.993 20160312_032600 0.802169 T F
20160521_040718 0.25,0.3,0.2 T 962.286 142.205 20160312_032600 0.833229 T F

04:10UT: TBad triggers and AdSec crash and does not recover

04:28UT: Recovered from AdSec crash

04:31UT: Collimating preset

04:38UT: Preset with Pyramid instead of QC

TN AO Gain AO CL TT Res. WF [nm rms] HO Res. WF [nm rms] AO rec DIMM ["] JCLSorted ascending PCL Notes
20160521_044418 0.85,0.35,0.25 T 151.016 126.692 20151212_233600 -1 T F
20160521_044429 0.85,0.35,0.25 T 160.429 123.73 20151212_233600 -1 T F
20160521_044439 0.85,0.35,0.25 T 175.543 147.399 20151212_233600 -1 T F

05:16UT: Mask can't be aligned. Preset with Pyramid instead of QC

No way of aligning this mask

05:25UT: AdSec Crash again

05:35UT: AdSec recovered

SDSSJ134332.85+415503

05:38UT: Collimating preset

05:42UT: Preset with Pyramid

Crash of PyrWfs SW

06:00UT: Closed loop

TN AO Gain AO CL TT Res. WF [nm rms] HO Res. WF [nm rms] AO rec DIMM ["] JCL PCL Notes
20160521_060218 0.85,0.3,0.25 T 129.272 137.082 20151212_233600 1.25581 T F
20160521_060231 0.85,0.3,0.25 T 133.426 131.507 20151212_233600 1.21283 T F
20160521_060241 0.85,0.3,0.25 T 121.353 130.923 20151212_233600 1.17607 T F

06:10UT: AdSec Crash again

Call Greg, removed actuator

06:30UT: Closed loop, mask aligning again

TN AO Gain AO CL TT Res. WF [nm rms] HO Res. WF [nm rms] AO rec DIMM ["] JCL PCL Notes
20160521_063539 0.85,0.3,0.25 T 128.135 122.794 20151212_233600 0.866286 T F
20160521_063550 0.85,0.3,0.25 T 125.237 137.464 20151212_233600 0.856651 T F
20160521_063559 0.85,0.3,0.25 T 131.518 144.467 20151212_233600 0.879053 T F

Hours of integrations without problems

08:20UT: Crash, spot big, low flux. We change target

SDSS1958 - test with DX

Switch to DX. Reconfigure telescope and edit scripts.

09:00UT: Collimation preset

Configured with pyramid and rec 20160520_010900. Loop crashes when enabling LO modes. Actually conditions are poor (seeing 1.6") and guiding with Jitter->PupilMirror seems to have a broken reconstructor (20160519_110910). Overall the SH is not well illuminated.

PseudoStaticModes configuration updated.

To measure Jitter->PupilMirror we should try in FLAO only closed loop. We created the CombinedRec 20160520_011900 from 20130917_214654 (was probably already available?)

LGSW Pupil motors in the position of CalSource: don't know if they are good on sky, probably yes.

JitterStabilizerInitialPosition is missing. What was a good one?