18:00: back to SX, we setup for the night

We start testing the new telescope offset procedure. We have a first try and the AOS hangs: Mar May 16 03:05:56

DX system worked fine, no hang, but PCS got any coord update.

Second test at 03:14:39

If we send a valid preset we do not hang the AOS...

22:00: fixed the software to acquire star by moving telescope we now use a new procedure: acquireRefAndNotifyTelescope

22:30: LB was right, we revert to his code

LgswUnitDx CL test

23:00: DX cal unit deployed, we start with calibrations.

We build combo recs w/ FLAO BIN2 and BIN4 for LGSW REC: 20160516_003800 (KL_V21):

  • FLAO BIN2: 20160303_190311 + LGSW REC: 20160516_003800 = COMBO TCoG48 + BIN2: 20160516_060600
  • FLAO BIN4: 20160303_182705 + LGSW REC: 20160516_003800 = COMBO TCoG48 + BIN2: 20160516_060700

Patrol Camera DX new target position

Beam PCX PCY
Blue 178 146
Yellow 170 113
Red 181 128
These new positions are saved into LaserAcquisitionTarget: 20160516_065700

00:40: closed on PYR BIN2 (MAG12), TS converged, no dist:

TN AO Gain AO CL TT Res. WF [nm rms] HO Res. WF [nm rms] AO rec DIMM ["] JCL PCL Notes
20160516_074138 1.5,0.5,0.5 T 91.688 97.1513 20160516_060600 -1 T F
20160516_074215 1.5,0.5,0.5 T 90.0031 97.5256 20160516_060600 -1 T F
20160516_074302 1,0.4,0.3 T 96.2551 86.7227 20160516_060600 -1 T F
20160516_074346 1,0.4,0.3 T 99.1832 85.6031 20160516_060600 -1 T F
The KL_v21 is really bad!

We switch to KL_v16, we calibrate for subaps: 20160312_213700

We generate a new set of MH:

KL # modes MH
16 10 20160516_080010
16 50 20160516_080050
16 150 20160516_080150
And then we acquire new IMs:

# modes IM REC LGSW IM LGSW REC COMBO REC
10 zero 20160516_080557 20160516_080900 -
50 20160516_080900 20160516_081155 20160516_081300 20160516_081400
2:00: we try to flatten the shell w/ FLAO BIN2 using KL_v16:

# modes IM REC LGSW IM LGSW REC COMBO REC
10 zero 20160516_090354 20160516_090900 -
50 20160516_090900 20160516_091126 20160516_091200 -
150 20160516_091200 20160516_091449 20160516_091200 -
150 20160516_091200 20160516_092003 20160516_092200 20160516_092300
150 20160516_092200 20160516_101111 20160516_101200 20160516_101300
Then we try to close the loop using 0.8" disturb:

TNSorted ascending AO Gain AO CL TT Res. WF [nm rms] HO Res. WF [nm rms] AO rec DIMM ["] JCL PCL Notes
20160516_100510 0.2,0.1,0.05 T 421.654 192.721 20160516_092300 -1 F F
20160516_100541 0.2,0.1,0.05 T 424.35 192.717 20160516_092300 -1 F F
20160516_100607 0.2,0.1,0.05 T 378.001 192.867 20160516_092300 -1 F F
20160516_100745 0.5,0.5,0.4 T 334.233 51.5399 20160516_092300 -1 F F
20160516_100813 0.5,0.5,0.4 T 332.125 50.9691 20160516_092300 -1 F F
20160516_100832 0.5,0.5,0.4 T 359.792 54.4035 20160516_092300 -1 F F
20160516_101522 0.5,0.5,0.4 T 204.078 55.1185 20160516_101300 -1 F F
20160516_101555 0.5,0.5,0.4 T 287.079 56.931 20160516_101300 -1 F F
20160516_101609 0.5,0.5,0.4 T 209.279 56.1518 20160516_101300 -1 F F

SDSS J1621+0607

3:30: we try to open, major HW failures: BCU dies, we can't close the loop

5:00: close