TN AO Gain Res. WF [nm rms] AO rec DIMM ["] JCL PCL Notes
20151219_001846 0, 0, 0 NA NA -1 T F PC trasmission test, HV off, PCT
20151219_001915 0, 0, 0 NA NA -1 T F PC trasmission test, 0.5 us, PCT
20151219_001934 0, 0, 0 NA NA -1 T F PC trasmission test, 1.0 us, PCT
20151219_001952 0, 0, 0 NA NA -1 T F PC trasmission test, 1.5 us, PCT
20151219_002010 0, 0, 0 NA NA -1 T F PC trasmission test, 2.0 us, PCT
20151219_001846_pcTransmission.jpg

18:00: lasers are launched.

19:00: end of collimation, heaters are on...

Hip9680

Magnitude 10 on catalog, we check first the LGS oscillation. It appears that no vibrations are present:

20151219_015900_patcams_spectra_withAndWithout.png

Then we optimize spot shape:

TN AO Gain Res. WF [nm rms] AO rec DIMM ["] JCL PCL Notes
20151219_020133 0, 0, 0 NA NA -1 F F LBFC opt
20151219_020150 0, 0, 0 NA NA -1 F F LBFC opt
20151219_020206 0, 0, 0 NA NA -1 F F LBFC opt
20151219_020222 0, 0, 0 NA NA -1 F F LBFC opt
20151219_020239 0, 0, 0 NA NA -1 F F LBFC opt
20151219_020256 0, 0, 0 NA NA -1 F F LBFC opt
20151219_020313 0, 0, 0 NA NA -1 F F LBFC opt
20151219_020329 0, 0, 0 NA NA -1 F F LBFC opt
20151219_020345 0, 0, 0 NA NA -1 F F LBFC opt
20151219_020402 0, 0, 0 NA NA -1 F F LBFC opt
20151219_020657 0, 0, 0 NA NA -1 F F LBFC opt
20151219_020713 0, 0, 0 NA NA -1 F F LBFC opt
20151219_020730 0, 0, 0 NA NA -1 F F LBFC opt
20151219_020746 0, 0, 0 NA NA -1 F F LBFC opt
20151219_020802 0, 0, 0 NA NA -1 F F LBFC opt
20151219_020818 0, 0, 0 NA NA -1 F F LBFC opt
20151219_020835 0, 0, 0 NA NA -1 F F LBFC opt
20151219_020851 0, 0, 0 NA NA -1 F F LBFC opt
20151219_020908 0, 0, 0 NA NA -1 F F LBFC opt
20151219_020923 0, 0, 0 NA NA -1 F F LBFC opt
20151229_021400_laser_spot_optimization.png

Maffei 1

19:30: we move to this target, magnitude 13 from catalog, 13 on PYR. OL FWHM is about 4.8px, CL is 1.5px. Dimm seeing ~0.6".

TN AO Gain Res. WF [nm rms] AO rec DIMM ["] JCL PCL Notes
20151219_022928 2,0.4,0.3 130.946 20151212_233600 -1 T F Auto
20151219_023040 2,0.4,0.3 133.83 20151212_233600 1.14874 T F Auto
20151219_023149 2,0.4,0.3 123.533 20151212_233600 1.17353 T F Auto
20151219_023300 2,0.4,0.3 113.615 20151212_233600 1.08072 T F Auto
20151219_023410 2,0.4,0.3 127.099 20151212_233600 1.06958 T F Auto
20151219_023519 2,0.4,0.3 125.554 20151212_233600 1.17519 T F Auto
20151219_023630 2,0.4,0.3 109.193 20151212_233600 1.08867 T F Auto
20151219_023740 2,0.4,0.3 119.886 20151212_233600 0.919463 T F Auto
20151219_025020 1.5,0.4,0.3 123.617 20151212_233600 1.12179 T F Auto
20151219_025131 1.5,0.45,0.35 134.469 20151212_233600 1.05883 T F Auto
20151219_025241 1.5,0.45,0.35 138.309 20151212_233600 1.12989 T F Auto
20151219_025351 1.5,0.45,0.35 721.531 20151212_233600 0.958839 T F Auto - Loop start to diverge
20151219_025701 1.5,0.45,0.35 454.61 20151212_233600 1.0042 F F Auto
20151219_025809 1.5,0.45,0.35 121.751 20151212_233600 0.927278 T F Auto
20151219_025919 1.5,0.45,0.35 127.319 20151212_233600 1.04901 T F Auto
20151219_030030 1.5,0.45,0.35 137.467 20151212_233600 1.03709 T F Auto
20151219_030140 1.5,0.45,0.35 126.96 20151212_233600 1.05911 T F Auto
20151219_030250 1.5,0.45,0.35 141.021 20151212_233600 0.806017 T F Auto
20151219_030400 1.5,0.45,0.35 219.395 20151212_233600 0.939912 T F Auto
20151219_030510 1.5,0.45,0.35 126.713 20151212_233600 0.85542 T F Auto
20151219_030620 1.5,0.45,0.35 225.349 20151212_233600 1.04549 T F Auto
20151219_031552 0,0.45,0.35 164.523 20151212_233600 -1 T F Auto
20151219_031702 2,0.45,0.35 118.094 20151212_233600 -1 T F Auto
20151219_031811 2,0.45,0.35 118.119 20151212_233600 -1 T F Auto
20151219_031921 2,0.45,0.35 114.688 20151212_233600 -1 T F Auto
20151219_032030 2,0.45,0.35 115.272 20151212_233600 -1 T F Auto
20151219_032138 2,0.45,0.35 109.695 20151212_233600 -1 T F Auto
20151219_032349 0.2,0.1,0.05 139.998 20151212_233600 -1 T F Auto
20151219_032458 2,0.45,0.35 113.697 20151212_233600 -1 T F Auto
20151219_032608 2,0.45,0.35 407.608 20151212_233600 -1 T F Auto
20151219_034251 0.6,0.45,0.35 133.702 20151212_233600 -1 T F Auto
20151219_034401 0,0.45,0.35 300.777 20151212_233600 -1 T F Auto
20151219_034512 0.6,0.45,0.35 142.404 20151212_233600 -1 T F Auto
20151219_034622 0.6,0.45,0.35 336.365 20151212_233600 -1 F F Auto - Jump
20151219_034730 0.6,0.45,0.35 315.347 20151212_233600 -1 T F Auto - No light on LGSW

NGC718

21:00: we change to this target. we guide on the galaxy core. magnitude 17 sent from preset, catalog ???, on Pyramid we measure 15.

TN AO Gain Res. WF [nm rms] AO rec DIMM ["] JCL PCL Notes
20151219_040740 0.6,0.45,0.35 833.799 20151212_233600 -1 T F Auto
20151219_040959 3.1,0.4,0.3 112.165 20151212_233600 -1 T F Auto
20151219_041109 0.2,0.1,0.05 88.672 20151212_233600 -1 T F Auto
20151219_041219 0.2,0.35,0.25 103.649 20151212_233600 -1 T F Auto
20151219_041329 0.2,0.1,0.05 99.8816 20151212_233600 -1 T F Auto
20151219_041439 0.2,0.1,0.05 103.645 20151212_233600 -1 T F Auto
20151219_041549 0.2,0.1,0.05 100.825 20151212_233600 0.719364 T F Auto
20151219_041700 0.2,0.1,0.05 98.821 20151212_233600 0.647282 T F Auto
20151219_041810 0.75,0.1,0.05 108.685 20151212_233600 0.728345 T F Auto
20151219_041921 0.75,0.1,0.05 101.486 20151212_233600 0.732451 T F Auto
20151219_042031 0, 0, 0 1898.77 20151212_233600 0.725999 T F Auto
20151219_043620 0.75,0.4,0.3 128.644 20151212_233600 0.712772 T F
20151219_043631 0.75,0.4,0.3 135.261 20151212_233600 0.733528 T F Auto
21:40: same target, new pointing after crash:

TN AO Gain Res. WF [nm rms] AO rec DIMM ["] JCL PCL Notes
20151219_044024 0,0.4,0.3 138.197 20151212_233600 0.719715 T F Auto
20151219_044134 0.75,0.4,0.3 118.705 20151212_233600 0.720627 T F Auto
20151219_044244 0.75,0.4,0.3 115.932 20151212_233600 0.680277 T F Auto
20151219_044354 0,0.4,0.3 4011.6 20151212_233600 0.649934 T F Auto
20151219_044503 0.75,0.4,0.3 116.333 20151212_233600 0.712211 T F Auto
20151219_044613 0.75,0.4,0.3 281.006 20151212_233600 0.715682 F F Auto
20151219_044721 0.75,0.4,0.3 161.73 20151212_233600 -1 T F Auto
20151219_044828 0.75,0.4,0.3 248.157 20151212_233600 -1 T F Auto
20151219_044937 0.75,0.4,0.3 452.31 20151212_233600 -1 T F Auto
20151219_045046 0.75,0.4,0.3 268.803 20151212_233600 -1 T F Auto
20151219_045154 0.75,0.4,0.3 393.547 20151212_233600 -1 T F Auto
20151219_045302 0.75,0.4,0.3 371.441 20151212_233600 -1 T F Auto
20151219_045410 0.75,0.4,0.3 429.991 20151212_233600 -1 F F Auto
20151219_045518 0.75,0.4,0.3 431.068 20151212_233600 -1 T F Auto
20151219_045627 0.75,0.4,0.3 458.191 20151212_233600 -1 T F Auto
20151219_045735 0.75,0.4,0.3 463.719 20151212_233600 -1 T F Auto
20151219_045843 0.75,0.4,0.3 532.125 20151212_233600 -1 T F Auto
20151219_045952 0.75,0.4,0.3 445.701 20151212_233600 -1 T F Auto
20151219_050100 0.75,0.4,0.3 467.638 20151212_233600 -1 T F Auto
20151219_050208 0.75,0.4,0.3 NA NA NA T F Auto
20151219_050325 0.2,0.1,0.05 175.955 20151212_233600 -1 T F Auto
20151219_050434 0.75,0.25,0.1 200.148 20151212_233600 -1 T F Auto
20151219_050545 0, 0, 0 703.93 20151212_233600 -1 T F Auto
20151219_050653 1.5,0.1,0.05 170.29 20151212_233600 -1 T F Auto
20151219_050804 1.5,0.1,1.38778e-17 161.342 20151212_233600 -1 T F Auto
20151219_050913 1.5,0.1,1.38778e-17 150.024 20151212_233600 -1 T F Auto
22:22: adsec restarted. number of modes applied to set flat: 0!

GAL192

23:30: we point to this target. From catalog we have magnitude 13.4, Pyramid measures 15.3.

TN AO Gain Res. WF [nm rms] AO rec DIMM ["] JCL PCL Notes
20151219_064633 1,0.2,0.1 160.605 20151212_233600 0.970933 T F Auto
20151219_064743 1.25,0.35,0.2 187.929 20151212_233600 1.12891 T F Auto
20151219_064852 1.25,0.35,0.2 510.677 20151212_233600 0.923288 F F Auto - bad AO loop
we're struggling in keeping the loop closed on LGSW, we move to Patrol cameras and check vibrations:

TN AO Gain Res. WF [nm rms] AO rec DIMM ["] JCL PCL Notes
20151219_072017 4.5,0.35,0.2 181.119 20151212_233600 1.01928 T F Auto
20151219_072127 4.5,0.25,0.05 220.695 20151212_233600 1.09605 T F Auto
20151219_072238 4.5,0.25,0.05 190.099 20151212_233600 1.08561 T F Auto
20151219_072448 0.1, 0, 0 1184.54 20151212_233600 1.21858 T F Auto
20151219_072558 4.5,0.25,1.38778e-17 180.996 20151212_233600 1.04224 T F Auto
20151219_072816 0,0.25,1.38778e-17 634.675 20151212_233600 1.12626 T F Auto - Bad

NGC2419

00:45: we point to EAST where there's less wind... Magnitude 13.4 from catalog, Pyramid reads 14.4. ~1" seeing.

TN AO Gain Res. WF [nm rms] AO rec DIMM ["] JCL PCL Notes
20151219_080259 0.7,0.4,0.25 142.31 20151212_233600 -1 T F Auto
20151219_080409 0.095,0.4,0.3 142.685 20151212_233600 0.866469 T F Auto
20151219_080519 1.5,0.4,0.3 116.158 20151212_233600 0.904923 T F Auto
20151219_080628 1.5, 0, 0 383.944 20151212_233600 0.903434 T F Auto
20151219_080738 1.5,0.4,0.3 131.321 20151212_233600 0.907625 T F Auto
20151219_080848 1.5,0.4,0.3 411.058 20151212_233600 -1 T F Auto
20151219_080957 1.5,0.4,0.3 364.185 20151212_233600 -1 T F Auto
20151219_081105 1.5,0.4,0.3 360.99 20151212_233600 -1 F F Auto
20151219_081213 1.5,0.4,0.3 374.001 20151212_233600 -1 T F Auto
20151219_081321 1.5,0.4,0.3 132.797 20151212_233600 -1 T F Auto
20151219_081432 1.5,0.4,0.3 149.7 20151212_233600 -1 T F Auto
20151219_081542 1.5,0.45,0.35 140.653 20151212_233600 0.928627 T F Auto
20151219_081651 1.5,0.45,0.35 135.237 20151212_233600 0.910928 T F Auto
20151219_081801 1.5,0.45,0.35 130.118 20151212_233600 0.89429 T F Auto
20151219_081910 1.5,0.45,0.35 180.27 20151212_233600 0.926095 T F Auto
20151219_082020 0.15,0.45,0.35 142.466 20151212_233600 0.988396 T F Auto
20151219_082131 1.5,0.45,0.35 146.402 20151212_233600 0.945749 T F Auto
20151219_082643 1.5,0.45,0.35 146.742 20151212_233600 -1 T F Auto
20151219_082754 1.5,0.45,0.35 140.579 20151212_233600 -1 T F Auto
20151219_082904 1.5,0.45,0.35 138.33 20151212_233600 -1 T F Auto
20151219_083014 0,0.45,0.35 2718.26 20151212_233600 -1 T F Auto
20151219_083124 1.5,0.45,0.35 142.713 20151212_233600 -1 T F Auto
20151219_083234 1.5,0.45,0.35 144.637 20151212_233600 -1 T F Auto
20151219_083344 1.5,0.45,0.35 164.582 20151212_233600 -1 T F Auto
20151219_083453 1.5,0.45,0.35 144.8 20151212_233600 -1 T F Auto
20151219_083603 1.5,0.45,0.35 332.936 20151212_233600 -1 F F Auto
20151219_084011 0,0.45,0.35 5455.8 20151212_233600 -1 T F Auto
20151219_084120 1.5,0.45,0.35 160.735 20151212_233600 -1 T F Auto
20151219_084231 1.5,0.45,0.35 177.301 20151212_233600 -1 T F Auto
20151219_084341 1.5,0.45,0.35 166.552 20151212_233600 -1 T F Auto
20151219_084451 1.5,0.45,0.35 168.716 20151212_233600 -1 T F Auto
20151219_084600 1.5,0.45,0.35 147.229 20151212_233600 -1 T F Auto
20151219_084711 1.5,0.45,0.35 152.714 20151212_233600 -1 T F Auto
20151219_084822 1.5,0.45,0.35 161.792 20151212_233600 -1 T F Auto
20151219_084932 1.5,0.45,0.35 168.629 20151212_233600 -1 T F Auto
20151219_085041 1.5,0.45,0.35 153.185 20151212_233600 -1 T F Auto
20151219_085152 1.5,0.45,0.35 150.873 20151212_233600 -1 T F Auto
20151219_085302 1.5,0.45,0.35 147.499 20151212_233600 -1 T F Auto
20151219_085412 1.5,0.45,0.35 144.217 20151212_233600 -1 T F Auto
20151219_085522 1.5,0.45,0.35 150.32 20151212_233600 -1 T F Auto
20151219_085632 1.5,0.45,0.35 155.973 20151212_233600 -1 T F Auto
20151219_085742 1.5,0.45,0.35 147.497 20151212_233600 -1 T F Auto
20151219_085852 1.5,0.45,0.35 152.887 20151212_233600 -1 T F Auto
20151219_090002 1.5,0.45,0.35 143.593 20151212_233600 -1 T F Auto
20151219_090111 1.5,0.45,0.35 145.576 20151212_233600 -1 T F Auto
20151219_090221 1.5,0.45,0.35 146.792 20151212_233600 -1 T F Auto
20151219_090331 1.5,0.45,0.35 155.701 20151212_233600 -1 T F Auto

VibComp test

2:00: we check for the VibComp system. Clouds are also appearing...

SDSSJ1110+6459

3:00: back to observation, we recollimate and point at this target. Magnitude 16 from catalog, we have magnitude 17 from pyramid.

We still have many clouds, red pupil is spilling from central subaps to side ones...

TN AO Gain Res. WF [nm rms] AO rec DIMM ["] JCL PCL Notes
20151219_101542 1,0.3,0.01 188.937 20151212_233600 0.733384 T F Auto
20151219_101652 1,0.3,0.01 181.499 20151212_233600 0.733384 T F Auto
20151219_101801 1,0.3,0.01 187.858 20151212_233600 0.733384 T F Auto
20151219_101911 1.1,0.3,0.01 158.403 20151212_233600 -1 T F Auto
20151219_102021 1.1,0.4,0.15 779.787 20151212_233600 -1 T F Auto
20151219_102132 1.1,0.4,0.01 236.209 20151212_233600 -1 T F Auto
20151219_102242 1.1,0.3,0.01 178.721 20151212_233600 1.06281 T F Auto
20151219_102352 1.1,0.3,0.01 151.755 20151212_233600 1.0424 T F Auto
20151219_102503 0.2,0.1,0.05 139.974 20151212_233600 1.05687 T F Auto
20151219_102612 1.1,0.3,0.01 198.024 20151212_233600 1.23282 T F Auto
20151219_102722 1.1,0.3,0.01 171.528 20151212_233600 1.06922 T F Auto
20151219_102832 1.1,0.3,0.01 680.168 20151212_233600 0.958191 T F Auto
20151219_102943 1.1,0.3,0.01 183.335 20151212_233600 1.06696 T F Auto
20151219_103053 1.1,0.3,0.01 198.581 20151212_233600 0.983 T F Auto
20151219_103204 1.1,0.3,0.01 456.346 20151212_233600 0.976139 T F Auto
20151219_103313 1.1,0.3,0.01 172.98 20151212_233600 1.08922 T F Auto
20151219_103422 1.1,0.3,0.01 208.117 20151212_233600 0.984021 T F Auto
20151219_103532 1.1,0.3,0.01 191.205 20151212_233600 1.05425 T F Auto

NGC4102

4:30: too many clouds! we point to this other target. We guide with TT on a double star.

20151219_115329 1.4,0.3, 0 247.911 20151212_233600 -1 T F Auto
20151219_115448 0,0.3, 0 505.737 20151212_233600 -1 F F Auto
20151219_115604 1.4,0.3,0.05 204.697 20151212_233600 -1 T F Auto
20151219_115723 1.4,0.3,0.05 219.953 20151212_233600 -1 T F Auto
20151219_115838 1.4,0.3,0.05 549.329 20151212_233600 -1 T F Auto
20151219_115950 1.4,0.3,0.05 460.013 20151212_233600 -1 T F Auto
20151219_120102 1.4,0.3,0.05 487.902 20151212_233600 -1 T F Auto
20151219_120216 1.2,0.3, 0 207.206 20151212_233600 -1 T F Auto
20151219_120331 1.35,0.45,0.35 200.999 20151212_233600 -1 T F Auto
20151219_120445 1.35,0.45,0.35 517.165 20151212_233600 -1 T F Auto
20151219_120559 0, 0, 0 560.084 20151212_233600 -1 T F Auto
20151219_120711 0, 0, 0 477.442 20151212_233600 -1 F F Auto
20151219_120856 0, 0, 0 714.215 20151212_233600 -1 F F Auto
20151219_121041 0, 0, 0 676.756 20151212_233600 -1 F F Auto
20151219_121220 0, 0, 0 626.854 20151212_233600 -1 F F Auto
20151219_121828 0, 0, 0 38.6651 20151212_233600 -1 F F Auto
20151219_121935 0, 0, 0 40.1799 20151212_233600 -1 F F Auto
20151219_122043 0, 0, 0 38.8871 20151212_233600 -1 F F Auto
20151219_122150 0, 0, 0 39.578 20151212_233600 -1 F F Auto
20151219_122258 0, 0, 0 38.7218 20151212_233600 -1 F F Auto
20151219_122406 0, 0, 0 38.9901 20151212_233600 -1 F F Auto
-- MarcoBonaglia - 11 Jan 2016
Topic revision: r7 - 28 Oct 2016, MarcoBonaglia
This site is powered by FoswikiCopyright © by the contributing authors. All material on this collaboration platform is the property of the contributing authors.
Ideas, requests, problems regarding Foswiki? Send feedback