Hip56107

19:45: new LGSW dark: 2050430_024510, we point Hip56107 sending the preset from LUCI

20:00: we fail... we send a preset from telescope

Seeing 0.9" from guider FWHM measurement

20:20: lasers are on-sky, click'n'go to acquire stars.

Focus and corrector plate scan

The focus stage scan moves between -100k and -160k steps. Then we move to -40k. The corrector plate is still at -1000 (rotator). We sweep the corrector between 100 and 800 steps.

Seeing 1.14"

The proper angle for rotator is 6500 steps, this is why we could not see differences in elongation...

We increase the limit for the corrector sweep between 50 and 950 steps

TN Notes Plot Plot (Threshold)
'20150430_032343','20150430_032417' Rotator @ 1000 20150430_032343_SpotEccentrVsLaunchCorrector.jpg 20150430_032343-20150430_032417_LGSThresholdVsBeamCorrector.jpg
'20150430_033030','20150430_033103' Rotator @ 6500 20150430_033030_SpotEccentrVsLaunchCorrector.jpg 20150430_033030-20150430_033103_LGSThresholdVsBeamCorrector.jpg
'20150430_033929','20150430_034009' Spot o PatCams 20150430_033929_SpotEccentrVsLaunchCorrector.jpg 20150430_033929-20150430_034009_LGSThresholdVsBeamCorrector.jpg
'20150430_034207','20150430_034250' Spot recentered on PatCams 20150430_034207_SpotEccentrVsLaunchCorrector.jpg 20150430_034207-20150430_034250_LGSThresholdVsBeamCorrector.jpg
Last scan was flat, so we defocus the stage up to -10k

TN Notes Plot Plot (Threshold)
'20150430_035458','20150430_035545' corrector scan between 100 and 900 steps, best is 600 20150430_035458_SpotEccentrVsLaunchCorrector.jpg 20150430_035458-20150430_035545_LGSThresholdVsBeamCorrector.jpg
Focus stage moved to -80k

TN Notes Plot Plot (Threshold)
'20150430_040248','20150430_040331' corrector scan between 100 and 900 steps, best is 600 20150430_040248_SpotEccentrVsLaunchCorrector.jpg 20150430_040248-20150430_040331_LGSThresholdVsBeamCorrector.jpg
Sweeping the corrector we found that it was not working

We try a focus scan:

TN Notes Plot (Threshold)
'20150430_041328','20150430_041419' focus scan between -180k and -40k steps, best is -120k 20150430_041328-20150430_041419_LGSThresholdVsFocusStage.jpg
21:00: JZ and SR go up to check corrector status

22:18: the corrector gear was dry, they put grease on it

We send a new LUCI preset, we fail again

22:23: lasers are back

Seeing 0.6" from DIMM

TN Notes Plot Plot (Threshold)
'20150430_052623','20150430_052704' corrector sweep, focus was -160k 20150430_052623_SpotEccentrVsLaunchCorrector.jpg 20150430_052623-20150430_052704_LGSThresholdVsBeamCorrector.jpg
'20150430_052934','20150430_053018'

corrector sweep, focus was -40k

Sweep movie: 20150430_052934_LBCsweep.mp4
20150430_052934_SpotEccentrVsLaunchCorrector.jpg 20150430_052934-20150430_053018_LGSThresholdVsBeamCorrector.jpg
'20150430_053723','20150430_053806' corrector sweep, focus was -100k 20150430_053723_SpotEccentrVsLaunchCorrector.jpg 20150430_053723-20150430_053806_LGSThresholdVsBeamCorrector.jpg
'20150430_054806','20150430_054845' corrector sweep, focus was -20k 20150430_054806_SpotEccentrVsLaunchCorrector.jpg 20150430_054806-20150430_054845_LGSThresholdVsBeamCorrector.jpg
'20150430_055156','20150430_055234' corrector sweep, focus was -20k 20150430_055156_SpotEccentrVsLaunchCorrector.jpg 20150430_055156-20150430_055234_LGSThresholdVsBeamCorrector.jpg

Vibration compensation test

23:00: Lasers on PatCams, snapshot are acquired

Focus and corrector sweep

TN Notes
'20150430_055823','20150430_055924' focus sweep, corrector was 900 (best # -150k)
'20150430_061548','20150430_061629' focus sweep, corrector was 500 (best # -130k)
'20150430_061717','20150430_061802' focus sweep, corrector was 200 (best # -130k)
'20150430_062015','20150430_062057' focus sweep, corrector was 900 (best # -150k)
We set the focus stage at -150k and we try a new swee of the corrector:

TN Notes Plot Plot (Threshold)
'20150430_062318','20150430_062356' corrector sweep, focus was -150k 20150430_062318_SpotEccentrVsLaunchCorrector.jpg 20150430_062318-20150430_062356_LGSThresholdVsBeamCorrector.jpg
Now we ut the light into the LGSW FoV and we measure the spot FW:

TN Notes
'20150430_063025' corrector 900, focus -150k, FW 1.55, II moment 3.48
'20150430_063051' corrector 900, focus -150k, FW 1.51, II moment 3.29
'20150430_063134' corrector 500, focus -150k, FW 1.57, II moment 3.55
'20150430_063147' corrector 500, focus -150k, FW 1.58, II moment 3.45
'20150430_063219' corrector 200, focus -150k, FW 1.61, II moment 3.98
'20150430_063234' corrector 200, focus -150k, FW 1.75, II moment 3.75
'20150430_063824' corrector 900, focus -150k, FW 1.43, II moment 3.17
Seeing is 1.0" from DIMM

TN AO rec LAN focus LM1 phi LM1 theta LAN state Bex B Bex R Bex Y LAS state JCL PM B PM R PM Y PCL LGSW state M2 hex Alt M3 Notes
20150430_070230 NA -180000 282.728 -85.2145 R2O 128500 124000 152000 R2O T 759,509 457,665 285,715 F ACQ -3.81401,-1.81606,1.78253,-308.533,-195.099,0.000748753 55.6122 -757.974,579.973,4.06799e-05,-25.9002
20150430_070311 NA -180000 282.728 -85.2145 R2O 128500 124000 152000 R2O T 759,509 457,665 285,715 F ACQ -3.67753,-1.78348,1.80203,-312.127,-181.079,-0.00012852 55.4697 -757.974,580.081, -0,-25.9002
20150430_070927 NA -180000 290.6 -89.7593 R2O 128500 124000 152000 R2O T 759,509 457,665 285,715 F ACQ -3.66675,-1.73806,1.7467,-318.811,-181.799,0.014486 54.1695 -757.974,579.973,4.06799e-05,-25.9002

Hip87681

00:20: We center the board on the QC tip: -76, -51.2, 45.0

We try to close the loop on the QC:

TN AO Gain Res. WF [nm rms] AO rec DIMM ["] JCL PCL Notes
20150430_072759 0.9,0.2,0.1 423.523 20150429_110600 0.910516 T F
20150430_073535 0.5,0.3,0.1 821.305 20150429_110600 0.976089 T F
20150430_081405 0, 0, 0 1077.31 20150429_110600 1.02998 T F
20150430_081432 0, 0, 0 1074.3 20150429_110600 1.10426 T F
20150430_081526 0,0.01, 0 947.465 20150429_110600 1.02697 T F
The ASM problem could be due to an ovetemperature of the electronics: we set the telescope back to Zenith and we go up to open the cabinet door. Suddently the temperature drops by 10°C.

2:00: board stages @ -74.5, -51.0, 55.0 nominal for LUCI operation when star is on-axis

Focus stage is set at -120k, elevation is 68°

TN AO Gain Res. WF [nm rms] AO rec DIMM ["] JCL PCL Notes
20150430_090425 0.1, 0, 0 1039.17 20150429_110600 1.09521 T F
20150430_090533 0.1, 0, 0 803.82 20150429_110600 0.992484 T F
20150430_090650 0.5, 0, 0 565.809 20150429_110600 1.06954 T F
20150430_090713 0.5, 0, 0 590.364 20150429_110600 1.10749 T F
Motors position on-sky, saved in new Pupil target: 20150501_044715

Pupil X Y
Blue 768 514
Yellow 270 722
Red 473 663
We are running at 650Hz anyhow the shell RIPs

TN AO Gain Res. WF [nm rms] AO rec DIMM ["] JCL PCL Notes
20150430_101530 0, 0, 0 1164.62 20150429_110600 -1 F F
20150430_101835 0, 0, 0 480.081 20150429_110600 -1 F F
20150430_102316 0, 0, 0 777.604 20150429_110600 0.938451 F F
20150430_102338 0, 0, 0 770.771 20150429_110600 0.88676 F F
20150430_102414 0, 0, 0 821.013 20150429_110600 0.941111 F F
20150430_103450 0, 0, 0 1400.49 20150429_110600 1.13008 F F
20150430_103520 0, 0, 0 1631.39 20150429_110600 1.32766 F F
20150430_103538 0, 0, 0 2393.22 20150429_110600 1.25763 F F
20150430_103555 0, 0, 0 2678.23 20150429_110600 1.23146 F F
20150430_103610 0, 0, 0 2968.74 20150429_110600 1.18221 F F
20150430_103627 0, 0, 0 2888.11 20150429_110600 1.13432 F F
20150430_103645 0, 0, 0 2470.16 20150429_110600 1.02111 F F
20150430_103659 0, 0, 0 2084.82 20150429_110600 0.975078 F F
20150430_103719 0, 0, 0 2040.14 20150429_110600 1.08423 F F
20150430_103735 0, 0, 0 2032.21 20150429_110600 1.12786 F F
20150430_103749 0, 0, 0 2070.82 20150429_110600 1.15036 F F
20150430_103804 0, 0, 0 1668.04 20150429_110600 1.23543 F F
20150430_104404 0, 0, 0 1666.09 20150429_110600 1.06028 F F
20150430_104424 0, 0, 0 1657.45 20150429_110600 0.797849 F F
20150430_104441 0, 0, 0 2086.75 20150429_110600 -1 F F
20150430_104456 0, 0, 0 1721.34 20150429_110600 -1 F F
20150430_104510 0, 0, 0 1650.88 20150429_110600 -1 F F
20150430_104526 0, 0, 0 1649 20150429_110600 -1 F F
20150430_104545 0, 0, 0 1663.59 20150429_110600 -1 F F
20150430_104601 0, 0, 0 1667.32 20150429_110600 -1 F F
20150430_104615 0, 0, 0 1756.55 20150429_110600 -1 F F
20150430_104633 0, 0, 0 2058.36 20150429_110600 -1 F F
20150430_104647 0, 0, 0 1793.44 20150429_110600 -1 F F
20150430_104701 0, 0, 0 1754.47 20150429_110600 -1 F F
20150430_104736 0, 0, 0 1731.9 20150429_110600 -1 F F
20150430_104755 0, 0, 0 1762.73 20150429_110600 -1 F F
20150430_104810 0, 0, 0 1701.36 20150429_110600 -1 F F
20150430_104826 0, 0, 0 1678.7 20150429_110600 -1 F F
20150430_104845 0, 0, 0 1672.14 20150429_110600 -1 F F
20150430_104906 0, 0, 0 1686.51 20150429_110600 -1 F F
20150430_104932 0, 0, 0 1519 20150429_110600 -1 F F
20150430_105010 0, 0, 0 1593.19 20150429_110600 -1 F F
20150430_105029 0, 0, 0 1632.22 20150429_110600 -1 F F
20150430_105044 0, 0, 0 1579.92 20150429_110600 -1 F F
20150430_105102 0, 0, 0 1554.36 20150429_110600 -1 F F
20150430_105118 0, 0, 0 1548.79 20150429_110600 -1 F F
20150430_105213 0, 0, 0 1556.64 20150429_110600 -1 F F
20150430_105238 0, 0, 0 1478.08 20150429_110600 -1 F F
20150430_105300 0, 0, 0 1631.52 20150429_110600 -1 F F
20150430_105320 0, 0, 0 1587.6 20150429_110600 -1 F F
20150430_105335 0, 0, 0 1629.92 20150429_110600 -1 F F
20150430_105351 0, 0, 0 1618.95 20150429_110600 -1 F F
20150430_105406 0, 0, 0 1642.74 20150429_110600 -1 F F
20150430_105420 0, 0, 0 1677.56 20150429_110600 -1 F F
20150430_105434 0, 0, 0 1622.61 20150429_110600 -1 F F
20150430_105450 0, 0, 0 1613.65 20150429_110600 -1 F F
20150430_105504 0, 0, 0 1612 20150429_110600 -1 F F
20150430_105519 0, 0, 0 1710.12 20150429_110600 -1 F F
20150430_105543 0, 0, 0 1661.81 20150429_110600 -1 F F
20150430_105557 0, 0, 0 1609.07 20150429_110600 -1 F F
20150430_105612 0, 0, 0 1613.77 20150429_110600 -1 F F
20150430_105655 0, 0, 0 1527.1 20150429_110600 -1 F F
20150430_105710 0, 0, 0 1592.18 20150429_110600 -1 F F
20150430_105734 0, 0, 0 1579.38 20150429_110600 -1 F F
20150430_105747 0, 0, 0 1604.63 20150429_110600 -1 F F
20150430_105803 0, 0, 0 1627.04 20150429_110600 -1 F F
20150430_105816 0, 0, 0 1732.01 20150429_110600 -1 F F
20150430_105830 0, 0, 0 1753.99 20150429_110600 -1 F F
20150430_105844 0, 0, 0 1748.59 20150429_110600 -1 F F
20150430_105858 0, 0, 0 1886.54 20150429_110600 -1 F F
20150430_105914 0, 0, 0 1867.81 20150429_110600 -1 F F
20150430_105927 0, 0, 0 1925.16 20150429_110600 -1 F F
20150430_105943 0, 0, 0 1966.71 20150429_110600 -1 F F
20150430_110012 0, 0, 0 1785.35 20150429_110600 -1 F F
20150430_110033 0, 0, 0 1732.49 20150429_110600 -1 F F
20150430_110156 0, 0, 0 2010.63 20150429_110600 -1 F F
20150430_110212 0, 0, 0 2049.12 20150429_110600 -1 F F
20150430_110236 0, 0, 0 2049.04 20150429_110600 -1 F F
20150430_110252 0, 0, 0 2302.15 20150429_110600 -1 F F
20150430_110310 0, 0, 0 2181.1 20150429_110600 -1 F F
20150430_110324 0, 0, 0 2797.43 20150429_110600 -1 F F
20150430_110340 0, 0, 0 2777.77 20150429_110600 -1 F F
20150430_110355 0, 0, 0 2923.39 20150429_110600 -1 F F
20150430_110409 0, 0, 0 2874.04 20150429_110600 -1 F F
20150430_110423 0, 0, 0 2862.1 20150429_110600 -1 F F
20150430_110437 0, 0, 0 3013.4 20150429_110600 -1 F F
20150430_110453 0, 0, 0 2916.15 20150429_110600 -1 F F
20150430_110539 0, 0, 0 2822.45 20150429_110600 -1 F F
20150430_110555 0, 0, 0 2306.83 20150429_110600 -1 F F
20150430_110609 0, 0, 0 2500.54 20150429_110600 -1 F F
20150430_110625 0, 0, 0 2436.59 20150429_110600 -1 F F
20150430_110638 0, 0, 0 2033.97 20150429_110600 -1 F F
20150430_110652 0, 0, 0 1942.5 20150429_110600 -1 F F
20150430_110706 0, 0, 0 2094.2 20150429_110600 -1 F F
20150430_110720 0, 0, 0 1764.81 20150429_110600 -1 F F
20150430_110733 0, 0, 0 1573.33 20150429_110600 -1 F F
20150430_110747 0, 0, 0 1810.9 20150429_110600 -1 F F
20150430_110800 0, 0, 0 1638.12 20150429_110600 -1 F F
20150430_110814 0, 0, 0 1614.73 20150429_110600 -1 F F
20150430_110849 0, 0, 0 1497.83 20150429_110600 -1 F F
20150430_110905 0, 0, 0 1464.91 20150429_110600 -1 F F
20150430_110919 0, 0, 0 1592.8 20150429_110600 -1 F F
20150430_110933 0, 0, 0 1411.54 20150429_110600 -1 F F
20150430_110947 0, 0, 0 1438.07 20150429_110600 -1 F F
-- MarcoBonaglia - 18 May 2015
I Attachment Action Size Date Who Comment
20150430_032343-20150430_032417_LGSThresholdVsBeamCorrector.jpgjpg 20150430_032343-20150430_032417_LGSThresholdVsBeamCorrector.jpg manage 49 K 29 Sep 2015 - 14:51 MarcoBonaglia  
20150430_032343_SpotEccentrVsLaunchCorrector.jpgjpg 20150430_032343_SpotEccentrVsLaunchCorrector.jpg manage 32 K 21 May 2015 - 16:03 MarcoBonaglia  
20150430_033030-20150430_033103_LGSThresholdVsBeamCorrector.jpgjpg 20150430_033030-20150430_033103_LGSThresholdVsBeamCorrector.jpg manage 50 K 29 Sep 2015 - 14:54 MarcoBonaglia  
20150430_033030_SpotEccentrVsLaunchCorrector.jpgjpg 20150430_033030_SpotEccentrVsLaunchCorrector.jpg manage 34 K 21 May 2015 - 16:01 MarcoBonaglia  
20150430_033929-20150430_034009_LGSThresholdVsBeamCorrector.jpgjpg 20150430_033929-20150430_034009_LGSThresholdVsBeamCorrector.jpg manage 43 K 29 Sep 2015 - 14:55 MarcoBonaglia  
20150430_033929_SpotEccentrVsLaunchCorrector.jpgjpg 20150430_033929_SpotEccentrVsLaunchCorrector.jpg manage 35 K 21 May 2015 - 16:01 MarcoBonaglia  
20150430_034207-20150430_034250_LGSThresholdVsBeamCorrector.jpgjpg 20150430_034207-20150430_034250_LGSThresholdVsBeamCorrector.jpg manage 42 K 29 Sep 2015 - 14:57 MarcoBonaglia  
20150430_034207_SpotEccentrVsLaunchCorrector.jpgjpg 20150430_034207_SpotEccentrVsLaunchCorrector.jpg manage 33 K 21 May 2015 - 16:02 MarcoBonaglia  
20150430_035458-20150430_035545_LGSThresholdVsBeamCorrector.jpgjpg 20150430_035458-20150430_035545_LGSThresholdVsBeamCorrector.jpg manage 42 K 29 Sep 2015 - 14:59 MarcoBonaglia  
20150430_035458_SpotEccentrVsLaunchCorrector.jpgjpg 20150430_035458_SpotEccentrVsLaunchCorrector.jpg manage 32 K 21 May 2015 - 15:01 MarcoBonaglia  
20150430_040248-20150430_040331_LGSThresholdVsBeamCorrector.jpgjpg 20150430_040248-20150430_040331_LGSThresholdVsBeamCorrector.jpg manage 46 K 29 Sep 2015 - 15:00 MarcoBonaglia  
20150430_040248_SpotEccentrVsLaunchCorrector.jpgjpg 20150430_040248_SpotEccentrVsLaunchCorrector.jpg manage 33 K 21 May 2015 - 15:07 MarcoBonaglia  
20150430_041328-20150430_041419_LGSThresholdVsFocusStage.jpgjpg 20150430_041328-20150430_041419_LGSThresholdVsFocusStage.jpg manage 42 K 29 Sep 2015 - 15:38 MarcoBonaglia  
20150430_052623-20150430_052704_LGSThresholdVsBeamCorrector.jpgjpg 20150430_052623-20150430_052704_LGSThresholdVsBeamCorrector.jpg manage 44 K 29 Sep 2015 - 15:02 MarcoBonaglia  
20150430_052623_SpotEccentrVsLaunchCorrector.jpgjpg 20150430_052623_SpotEccentrVsLaunchCorrector.jpg manage 38 K 21 May 2015 - 15:13 MarcoBonaglia  
20150430_052934-20150430_053018_LGSThresholdVsBeamCorrector.jpgjpg 20150430_052934-20150430_053018_LGSThresholdVsBeamCorrector.jpg manage 46 K 29 Sep 2015 - 15:03 MarcoBonaglia  
20150430_052934_LBCsweep.mp4mp4 20150430_052934_LBCsweep.mp4 manage 165 K 29 Sep 2015 - 17:39 MarcoBonaglia  
20150430_052934_SpotEccentrVsLaunchCorrector.jpgjpg 20150430_052934_SpotEccentrVsLaunchCorrector.jpg manage 34 K 21 May 2015 - 15:15 MarcoBonaglia  
20150430_053723-20150430_053806_LGSThresholdVsBeamCorrector.jpgjpg 20150430_053723-20150430_053806_LGSThresholdVsBeamCorrector.jpg manage 45 K 29 Sep 2015 - 15:51 MarcoBonaglia  
20150430_053723_SpotEccentrVsLaunchCorrector.jpgjpg 20150430_053723_SpotEccentrVsLaunchCorrector.jpg manage 37 K 21 May 2015 - 15:18 MarcoBonaglia  
20150430_054806-20150430_054845_LGSThresholdVsBeamCorrector.jpgjpg 20150430_054806-20150430_054845_LGSThresholdVsBeamCorrector.jpg manage 42 K 29 Sep 2015 - 15:06 MarcoBonaglia  
20150430_054806_SpotEccentrVsLaunchCorrector.jpgjpg 20150430_054806_SpotEccentrVsLaunchCorrector.jpg manage 33 K 21 May 2015 - 15:22 MarcoBonaglia  
20150430_055156-20150430_055234_LGSThresholdVsBeamCorrector.jpgjpg 20150430_055156-20150430_055234_LGSThresholdVsBeamCorrector.jpg manage 43 K 29 Sep 2015 - 15:08 MarcoBonaglia  
20150430_055156_SpotEccentrVsLaunchCorrector.jpgjpg 20150430_055156_SpotEccentrVsLaunchCorrector.jpg manage 33 K 21 May 2015 - 15:23 MarcoBonaglia  
20150430_062318-20150430_062356_LGSThresholdVsBeamCorrector.jpgjpg 20150430_062318-20150430_062356_LGSThresholdVsBeamCorrector.jpg manage 45 K 29 Sep 2015 - 15:09 MarcoBonaglia  
20150430_062318_SpotEccentrVsLaunchCorrector.jpgjpg 20150430_062318_SpotEccentrVsLaunchCorrector.jpg manage 37 K 21 May 2015 - 15:26 MarcoBonaglia  
Topic revision: r4 - 29 Sep 2015, MarcoBonaglia
This site is powered by FoswikiCopyright © by the contributing authors. All material on this collaboration platform is the property of the contributing authors.
Ideas, requests, problems regarding Foswiki? Send feedback