Fast guiding check w/ oscilloscope

11:30: Tommaso and Sebastian go on the telescope to check the Fast Guiding delay.

We connect an oscilloscope after the 10x divisor installed by Julian Z. The output of the BCU is in the range 0-100V, with the 10x divisor the signal to the NIbox is in the range 0-10V. This image shows the electrical scheme of the 10x divisor and the frequency transfer function.

bcu_to_ni_sim.jpg

In the first test we send two step offset with the ARGOS software. The step are 0.005 rad, every 0.1 seconds. We notice that the signals are double steps, the first is 1V and the second is remnant part of the signal.

IMG_20160515_120156.jpg

We connect the output of the PI module to the oscilloscope and we check the delay. The PI signal has a delay of ~1ms, this is consistent with NI box delay measured by Diethard in MPIA. The PI control is ON. The settling time is 4ms.

IMG_20160515_121021.jpg

We decrease the steps, now are 0.001 rad, every 0.1 seconds. The double step disappear. The settling time now seems be 3ms.

IMG_20160515_121923.jpg

We connect the signal form the BCU (with the 10x divisor) directly to the PI module and we check the signals. We notice that the output signal has no delay.

IMG_20160515_122340.jpg

14:30: Tommaso and Sebastian go back on the telescope to make some other measure and install the WfsCameraPump1 on the pnCCD of the WfsCamera1. Pressure inside the pnCCD was 12mbar.

0.002 rad step with 10x divisor

IMG_20160515_143458.jpg

0.004 rad step with 10x divisor

IMG_20160515_143521.jpg

0.008 rad step with 10x divisor

IMG_20160515_143730.jpg

0.004 rad step without 10x divisor

IMG_20160515_144157.jpg

0.008 rad step on RED channel. We disable the control loop and we move the RED mirror with the "argos.hvc.channel" command "moveTo"

IMG_20160515_145610.jpg

0.008 rad step on GREEN channel with the "argos.hvc.channel" command "moveTo"

IMG_20160515_172300.jpg

17:30: The pressure in the pnCCD on the SX side was 0.34mbad. We unmount the pump.

LgswUnitDx calibration and CL test

We setup for calibrations on DX-LGSW.

DX HEXA -5.45 -8.17 0.87 252 82
DX BOARD -72.6 -56.8 52.8

Pupil PMX PMY
Blue 410 550
Yellow 290 390
Red 425 365

We can launch w/ RO offsets: -0.5; -6.5mm

  • DX and SX RO X coordinates are opposite: must be changed in the AARB

We setup for Pyramid BIN4:

TN AO Gain AO CL TT Res. WF [nm rms] HO Res. WF [nm rms] AO rec DIMM ["] JCL PCL Notes
20160515_200314 2.2,0.4,0.3 T 105.371 37.6585 20150704_021500 -1 T F
20160515_200750 2.2,0.4,0.3 T 124.694 37.1453 20150704_021500 -1 T F

14:00: we try to acquire new LGSW IM, subaps: 20160312_213700

Pupil PMX PMY
Blue 410 550
Yellow 290 390
Red 415 365

Pupil position saved as tag: 20160515_210710

We generate new modal histories:

KL # modes MH
21 10 20150515_220100
21 50 20150515_220500
21 150 20150515_221500

We start calibrating 10 modes:

# modes IM REC LGSW IM LGSW REC COMBO REC
10 zero 20160515_233733 20160515_234300 20160515_234900
50 zero 20160516_002111 20160516_002400 -
50 20160515_234300 20160516_002648 20160516_003300 -
150 20160515_234300 20160516_003609 20160516_003800 20160516_003900

DX HEXA -5.66 -9.34 0.88 364 58

Then we load combo REC: 20160516_003900

TN AO Gain AO CL TT Res. WF [nm rms] HO Res. WF [nm rms] AO rec DIMM ["] JCL PCL Notes
20160516_004017 0.5,0.5,0.5 T 247.779 68.2895 20160516_003900 -1 F F
20160516_004047 0.5,0.5,0.5 T 264.184 68.0645 20160516_003900 -1 F F
20160516_004201 0.5,0.5,0.5 T 353.978 99.9707 20160516_003900 -1 F F
20160516_004228 0.5,0.5,0.5 T 255.783 97.4896 20160516_003900 -1 F F
20160516_004254 0.5,0.5,0.5 T 251.248 97.7947 20160516_003900 -1 F F

After the test we have a new hexapod position:

DX HEXA -5.46 -7.87 0.88 211 78
I Attachment Action Size Date Who Comment
20160516_old_KLv16_new_KLv21_noDist.pngpng 20160516_old_KLv16_new_KLv21_noDist.png manage 136 K 16 May 2016 - 23:31 UnknownUser  
20160516_old_and_new_KLv16_08dist.pngpng 20160516_old_and_new_KLv16_08dist.png manage 140 K 16 May 2016 - 23:31 UnknownUser  
IMG_20160515_120156.jpgjpg IMG_20160515_120156.jpg manage 2 MB 15 May 2016 - 19:47 UnknownUser  
IMG_20160515_120802.jpgjpg IMG_20160515_120802.jpg manage 2 MB 15 May 2016 - 19:47 UnknownUser  
IMG_20160515_121021.jpgjpg IMG_20160515_121021.jpg manage 2 MB 15 May 2016 - 19:47 UnknownUser  
IMG_20160515_121923.jpgjpg IMG_20160515_121923.jpg manage 1 MB 15 May 2016 - 19:47 UnknownUser  
IMG_20160515_122340.jpgjpg IMG_20160515_122340.jpg manage 1004 K 15 May 2016 - 19:47 UnknownUser  
IMG_20160515_143458.jpgjpg IMG_20160515_143458.jpg manage 1 MB 15 May 2016 - 22:13 UnknownUser  
IMG_20160515_143521.jpgjpg IMG_20160515_143521.jpg manage 1 MB 15 May 2016 - 22:18 UnknownUser  
IMG_20160515_143730.jpgjpg IMG_20160515_143730.jpg manage 1 MB 15 May 2016 - 22:13 UnknownUser  
IMG_20160515_143733.jpgjpg IMG_20160515_143733.jpg manage 1 MB 15 May 2016 - 22:13 UnknownUser  
IMG_20160515_144157.jpgjpg IMG_20160515_144157.jpg manage 1 MB 15 May 2016 - 22:13 UnknownUser  
IMG_20160515_145521.jpgjpg IMG_20160515_145521.jpg manage 1 MB 15 May 2016 - 22:13 UnknownUser  
IMG_20160515_145610.jpgjpg IMG_20160515_145610.jpg manage 1 MB 15 May 2016 - 22:13 UnknownUser  
IMG_20160515_172300.jpgjpg IMG_20160515_172300.jpg manage 1 MB 16 May 2016 - 00:51 UnknownUser  
bcu_to_ni_sim.jpgjpg bcu_to_ni_sim.jpg manage 543 K 15 May 2016 - 19:47 UnknownUser  
mode35_KLv16.pngpng mode35_KLv16.png manage 85 K 16 May 2016 - 23:31 UnknownUser  
Topic revision: r8 - 24 May 2016, MarcoBonaglia
This site is powered by FoswikiCopyright © by the contributing authors. All material on this collaboration platform is the property of the contributing authors.
Ideas, requests, problems regarding Foswiki? Send feedback