Servicing of LgswUnitDx

Entance windows separation is 137mm (nominal is 137.9mm).

IMG_20141030_161250651.jpg

This is the same distance that we have in the LgswUnitSx. So the field stop is actually at

Buco nella shield delle pockels cell per pompare aria. Festo 6mm Accorciati i cavi high voltage.

Put plastic wrap around hv cables to increase separation.

PockelsCell2 = Blue

IMG_view_blue_cell.jpg

PockelsCell3 = Yellow

IMG_view_yellow_cell.jpg

PockelsCell4 = Red

IMG_view_red_cell.jpg

Hv calbes inside the EM box show some king of white deposition (very similar to cotton stuff), we have cleaned it with alchool.

IMG_20141030_145711163.jpg

Cambiata flangia input.

Cambiate entrance windows.

Cambiata calibration unit.

Check light with auxiliary Thorlabs diode: is ok on internal source.

Reinstalled old patrol cameras.

-- LorenzoBusoni - 31 Oct 2014

I AttachmentSorted ascending Action Size Date Who Comment
IMG_20141030_145711163.jpgjpg IMG_20141030_145711163.jpg manage 1002 K 31 Oct 2014 - 21:30 MarcoBonaglia  
IMG_20141030_161250651.jpgjpg IMG_20141030_161250651.jpg manage 1 MB 31 Oct 2014 - 21:27 MarcoBonaglia  
IMG_view_blue_cell.jpgjpg IMG_view_blue_cell.jpg manage 534 K 04 Nov 2014 - 03:22 MarcoBonaglia  
IMG_view_red_cell.jpgjpg IMG_view_red_cell.jpg manage 630 K 04 Nov 2014 - 03:24 MarcoBonaglia  
IMG_view_yellow_cell.jpgjpg IMG_view_yellow_cell.jpg manage 824 K 04 Nov 2014 - 03:23 MarcoBonaglia  
This topic: ARGOS > WebHome > Commissioning > DiaryRun3 > DayTime20141030
Topic revision: 04 Nov 2014, MarcoBonaglia
This site is powered by FoswikiCopyright © by the contributing authors. All material on this collaboration platform is the property of the contributing authors.
Ideas, requests, problems regarding Foswiki? Send feedback