Known history

When Where Who What
201204xx OAA   BCU used by HD SW group to develop the software is shipped to Arcetri
20120412 OAA OAA IP address changed to 192.168.18.30
201212xx OAA OAA IP address changed to 192.168.31.141
20130215     goes back to Munich together with Julian and Gilles after DX rack delivery
20130610 OAA OAA installed on LgswRack2 and IP changed to 192.168.32.140
20130710 OAA OAA after lighting doesn't communicate, sent to microgate
20130718 OAA OAA Microgate to OAA with new Ethernet board. MAC address changed from 00:20:af:f4:cc:34 to 00:20:af:f4:cc:54
20130730 OAA OAA Since HVC board #1 continues to fail MB removes this BCU from LgswRack2
20130930     goes back to Munich to Gilles
20140203 OAA MA Fixed by Microgate for fastlink issue
20140204 OAA LB BCU Logic updated to 11.70. Current IP is 192.168.31.141
20140403 OAA MA, GOX, LB BCU Logic updated to 11.71. Current IP is 192.168.32.141 ?
20140415 OAA MA, GOX, LB BCU Logic updated to 11.71 (fixed). DSP Logic updated to 8.46
20140415 OAA LB Installed on LgswRack2 and configured as SxBcu
20140505 OAA TM Removed HVC board 0008 (channel 0 broken) and shipped to MPE
20140516 OAA MB Installed HVC board 10 received from MPE
20140703 OAA MB, GOX NIOS 5.09.0046 installed to fix param selector issue.
20141022 OAA MB Temporarly swapped HVC board 010 with 008 on DSP02 to check output voltages. After test HVC 010 is reinstalled.
20150515 OAA MB, LB Power supply voltage raised from 3.31V to 3.48V
20150527 OAA MB Swapped HVC board 0007 with 0010 to check for setup failures
20150605 OAA MB Installed HVC board 0003 on DSP2, HVC board 0007 will be shipped to MicrogateSrl
20160313 LBT GOX, JZ Power supply voltage raised from 3.44V to 3.49V
20160513 LBT TM, MB Removed from LgswRack2
20160514 LBT JZ, MB Power supply voltage trimmed to 3.65V. HVC#0010 installed in DSP4. Installed in LgswRack2.
20160525 LBT GR, LB Removed from LgswRack2. Ready to be shipped to MicrogateSrl
201608xx MIC MA Modification on the minicrate power supply to improve the long term stability of the output voltage. With the modification the voltage level on the boards is now very stable and, considering the adopted solution, we assure this stability over long time. (As from MA email of 7/9/2015)
20160925 MPE SR Received from MicrogateSrl and shipped to LBT
20161005 LBT GR Received from MPE and stored at LBTLevel3L in a gray aluminum box
20161012 LBT MB, WG Installed on LgswRack2
20200921 LBT GR Removed from SX LgswRack2 due to failure to Set Up and stored in ArgosCabinet0 (until it is decided if needs to be sent to Microgate). Replaced with BcuNumber3.
20210622 LBT GR Sent to MicrogateSrl for repair.
20220301 LBT GR Received from MicrogateSrl.
20220305 LBT GR Taken to the observatory. Stored in LBTLevel3U, ArgosCabinet3

Known issues

This unit historically had issues with ethernet communication and HVC configuration.

When What
20130617 several ETH failures, board dismounted, screws tightened and reassembled...
20140501 (LB/AP) the HVC board 0008 has the first channel broken after testing the transfer function with time history
20150304 ThingsThatIDontUnderstand0002 happened 5 times in a row...
20150528 After swapping HVC boards 0010 and 0007 the error ThingsThatIDontUnderstand0002 follows board 0007, see LB email to Mario
20160513 In the last 2 days it continued to hung after 2-5min of CL operation (mainly w/ PyrWfs as TTS). Removed from LgswRack2, HVC#0010 moved to BcuNumber2

Current known config

Crate ID Crate MAC Crate IP
0x00 00:20:af:f4:cc:54 192.168.32.140

Board S/N FPGA NIOS Notes
BCU 00068 v11.71 v5.09.0046  
DSP00 1111 v8.46 v5.06.0040 ADC
DSP02 0003 v2.02 v4.02.0020 hvc
DSP04 0010 v2.02 v4.02.0020 hvc

Previous known config

Until 20160513:

Crate ID Crate MAC Crate IP
0x00 00:20:af:f4:cc:54 192.168.32.140

Board S/N FPGA NIOS Notes
BCU 00068 v11.71 v5.09.0046  
DSP00 1111 v8.46 v5.06.0040 ADC
DSP02 0003 v2.02 v4.02.0020 hvc
DSP04 0007 v2.02 v4.02.0020 hvc
Until 05/06/2015:

Board S/N FPGA NIOS Notes
BCU 00068 v11.71 v5.09.0046  
DSP00 1111 v8.46 v5.06.0040 ADC
DSP02 0007 v2.02 v4.02.0020 hvc
DSP04 0010 v2.02 v4.02.0020 hvc
Until 27/05/2015 (HVC board swap):

Board S/N FPGA NIOS Notes
BCU 00068 v11.71 v5.09.0046  
DSP00 1111 v8.46 v5.06.0040 ADC
DSP02 0010 v2.02 v4.02.0020 hvc
DSP04 0007 v2.02 v4.02.0020 hvc
Until 03 Jul 2014

Crate ID Crate MAC Crate IP
0x00 00:20:af:f4:cc:54 192.168.32.140

Board S/N FPGA NIOS Notes
BCU 00068 v11.71 v5.08.0046  
DSP00 1111 v8.46 v5.06.0040 ADC
DSP02 0010 v2.02 v4.02.0020 hvc
DSP04 0007 v2.02 v4.02.0020 hvc
Until 16 Jun 2014

Crate ID Crate MAC Crate IP
0x00 00:20:af:f4:cc:54 192.168.32.140

Board S/N FPGA NIOS Notes
BCU 00068 v11.71 v5.08.0046  
DSP00 1111 v8.46 v5.06.0040 ADC
DSP02 0008 v2.02 v4.02.0020 hvc
DSP04 0007 v2.02 v4.02.0020 hvc
Until 15 Apr 2014

Crate ID Crate MAC Crate IP
0x00 00:20:af:f4:cc:54 192.168.32.140

Board S/N FPGA NIOS Notes
BCU 00068 v11.71 v5.08.0046  
DSP00 1111 v8.45 v5.06.0040 ADC
DSP02 0008 v2.02 v4.02.0020 hvc
DSP04 0007 v2.02 v4.02.0020 hvc
Until 02 Apr 2014

Crate ID Crate MAC Crate IP
0x00 00:20:af:f4:cc:54 192.168.31.141

Board S/N FPGA NIOS Notes
BCU 00068 v11.70 v5.08.0046  
DSP00 1111 v8.45 v5.06.0040 ADC
DSP02 0008 v2.02 v4.02.0020 hvc
DSP04 0007 v2.02 v4.02.0020 hvc
Until 03 Feb 2014

Crate ID Crate MAC Crate IP
0x00 00:20:af:f4:cc:54 192.168.31.141

Board S/N FPGA NIOS Notes
BCU 00068 v11.68 v5.07.0046  
DSP00 1111 v8.45 v5.06.0040 ADC
DSP02 0008 v2.01 v4.02.0020 hvc
DSP04 0007 v2.01 v4.02.0020 hvc

Configuration screenshots

obtained with EngPanel:

2012/04/12 (LB): BCU config screenshot, BCU status screenshot, DSP#0 screenshot, DSP#2 screenshot, DSP#4 screenshot

2013/07/10 (MB): BCU config screenshot, BCU status screenshot, DSP#0 screenshot, DSP#2 screenshot, DSP#4 screenshot

Maintenance

See BCU and EngPanel

List of components devices

DeviceInfoForm edit

Name BCU #1
SerialNumber 0001
Location ArgosCabinet3
MACAddress 00:20:af:f4:cc:54
DeviceType BCU
Service Tag
I Attachment Action Size Date Who Comment
BCU_config.jpgjpg BCU_config.jpg manage 326 K 10 Jul 2013 - 09:28 MarcoBonaglia  
BCU_status.jpgjpg BCU_status.jpg manage 303 K 10 Jul 2013 - 09:31 MarcoBonaglia  
DSP#00.jpgjpg DSP#00.jpg manage 415 K 10 Jul 2013 - 09:31 MarcoBonaglia  
DSP#02.jpgjpg DSP#02.jpg manage 406 K 10 Jul 2013 - 09:31 MarcoBonaglia  
DSP#04.jpgjpg DSP#04.jpg manage 406 K 10 Jul 2013 - 09:31 MarcoBonaglia  
DSP00.jpgjpg DSP00.jpg manage 415 K 10 Jul 2013 - 09:36 MarcoBonaglia  
DSP02.jpgjpg DSP02.jpg manage 406 K 10 Jul 2013 - 09:36 MarcoBonaglia  
DSP04.jpgjpg DSP04.jpg manage 406 K 10 Jul 2013 - 09:36 MarcoBonaglia  
Diapositiva1.JPGJPG Diapositiva1.JPG manage 98 K 12 Apr 2012 - 15:51 LorenzoBusoni BCU config screenshot
Diapositiva2.JPGJPG Diapositiva2.JPG manage 75 K 12 Apr 2012 - 15:51 LorenzoBusoni BCU status screenshot
Diapositiva3.JPGJPG Diapositiva3.JPG manage 125 K 12 Apr 2012 - 15:52 LorenzoBusoni DSP0 screenshot
Diapositiva4.JPGJPG Diapositiva4.JPG manage 118 K 12 Apr 2012 - 15:52 LorenzoBusoni DSP1 screenshot
Diapositiva5.JPGJPG Diapositiva5.JPG manage 108 K 12 Apr 2012 - 15:53 LorenzoBusoni DSP4 screenshot
This topic: ARGOS > WebHome > ArgosDeviceList > BCU > BcuNumber1
Topic revision: 08 Mar 2022, GustavoRahmer
This site is powered by FoswikiCopyright © by the contributing authors. All material on this collaboration platform is the property of the contributing authors.
Ideas, requests, problems regarding Foswiki? Send feedback